The Role of Artificial Neural Networks in Prostate Magnetic Resonance Imaging (MRI) Segmentation
Cureus · 29.07.2026
DOI: 10.7759/cureus.113599 ↗Kurz gefasst
Fachartikel in der Zeitschrift Cureus, erschienen am 29.07.2026. Verfasst von Puranjay Shori und 5 weiteren Autor:innen. Beteiligte Einrichtung: Medicine, University of Texas (UT) Southwestern Medical School, Dallas, USA. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Introduction
In recent years, artificial intelligence (AI)-generated prostate whole-gland segmentation has shown promise for clinical use. This study compares AI- and human-derived prostate segmentation on magnetic resonance imaging (MRI) and evaluates its practical application.
Methods
A retrospective study from December 2020 to December 2022 evaluated 31 randomly selected patients who had previously undergone MRI-ultrasound (US)-guided fusion biopsies. AI-generated auto-contours of the whole-gland prostate, seminal vesicle, and urethra were produced using the ProtégéAI feature of the MIM Software from MRIs obtained on GE Signa HDxt 3.0T and Siemens Altea Magnetom 1.5T scanners. The same MRIs were independently contoured manually by a board-certified urologist (U) and a board-certified radiologist (R) using MIM Software contouring tools. Volumetric conformity among the three contour sets was assessed using the Dice similarity coefficient, Hausdorff distance (HD), and mean distance to agreement (MDA). Contouring time was recorded for each method, with AI timed from command execution to file generation and physicians timed from file opening to save. Pairwise Wilcoxon signed-rank tests (JMP Pro 15) compared contouring times and similarities across AI to urologist (AI-U), AI to radiologist (AI-R), and urologist to radiologist (U-R) contours.
Results
The average volumetric Dice similarity, HD, and MDA for the AI-U contours were 0.875±0.039, 9.186±4.004 mm, and 1.410±0.511 mm for the whole-gland prostate, 0.283±0.185, 18.117±13.265 mm, and 4.907±4.992 mm for the urethra, and 0.377±0.244, 14.708±9.489 mm, and 4.260±4.092 mm for the seminal vesicle. For the AI-R contours, the values were 0.757±0.072, 17.562±7.240 mm, and 3.050±1.339 mm for the prostate, 0.162±0.105, 19.956±11.962 mm, and 4.798±3.879 mm for the urethra, and 0.451±0.219, 16.069±9.245 mm, and 4.075±3.734 mm for the seminal vesicle. For the U-R contours, the values were 0.769±0.070, 15.109±6.177 mm, and 2.733±1.265 mm for the prostate, 0.144±0.091, 13.560±7.087 mm, and 3.406±1.985 mm for the urethra, and 0.471±0.216, 12.981±6.756 mm, and 3.264±2.308 mm for the seminal vesicle. Using the Wilcoxon signed-rank test with statistical significance defined as p
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Medicine, University of Texas (UT) Southwestern Medical School, Dallas, USA.
- Urology, University of Florida, Gainesville, USA.
- Biology and Psychology, University of Florida, Gainesville, USA.
- MRI Physics, Cortechs.ai, Orlando, USA.
- Radiology, HCA Florida North Florida Hospital, Gainesville, USA.
- Urology, National Medical Association and Research Group, Gainesville, USA.
Zitieren
APA
Shori, P., Varela, J., Shah, S., Amin, M., Kim, J., & Bird, V. (2026). The Role of Artificial Neural Networks in Prostate Magnetic Resonance Imaging (MRI) Segmentation. Cureus. https://doi.org/10.7759/cureus.113599
Vancouver
Shori P, Varela J, Shah S, Amin M, Kim J, Bird V. The Role of Artificial Neural Networks in Prostate Magnetic Resonance Imaging (MRI) Segmentation. Cureus. 2026. doi:10.7759/cureus.113599
BibTeX
@article{Shori2026,
author = {Shori, Puranjay and Varela, Juan and Shah, Shreya and Amin, Manish and Kim, Jong and Bird, Victoria},
title = {The Role of Artificial Neural Networks in Prostate Magnetic Resonance Imaging (MRI) Segmentation},
journal = {Cureus},
year = {2026},
doi = {10.7759/cureus.113599}
}
EUVIMED