EUVIMEDEUVIMED entdecken
Crossref

Diagnostic Accuracy of a Novel Fluorescent Immunoassay-Based Test in Latent Tuberculosis Infection

Ayse N Ceylan, Harbiye D Canat, Selda Komec, Abdurrahman Gulmez

Clinical Laboratory · 2025

DOI: 10.7754/clin.lab.2024.241003 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Laboratory, 2025. Verfasst von Ayse N Ceylan und 3 weiteren Autor:innen. Hauptthemen laut MEDLINE: Latent Tuberculosis, Interferon-gamma Release Tests, Mycobacterium tuberculosis, Fluorescent Antibody Technique.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Tuberculosis is still a significant global health challenge, with latent tuberculosis infection (LTBI) posing a risk for the development of active disease. This study aimed to compare the performance of the STAN-DARD F TB-Feron FIA test with the QuantiFERON-TB Gold Plus, a test endorsed by the World Health Organization (WHO), in detecting LTBI. METHODS: We included 137 participants, out of which 58.4% were healthy and 41.6% were undergoing immunosuppressive treatment. RESULTS: The agreement between QFT-Plus and TB-Feron FIA was substantial, with an overall concordance of 83.0% and a Cohen's kappa coefficient of 0.631 (p < 0.001). The TB-Feron FIA test demonstrated a sensitivity of 84.38% and a specificity of 87.10% compared to QFT-Plus. When TB-Feron FIA cutoff values were recalculated according to QFT Plus results, the value of 0.37 gave the highest sensitivity and specificity (84.37% and 88.42%). CONCLUSIONS: Our findings suggest that TB-Feron FIA is a reliable alternative for diagnosing LTBI, with performance comparable to QFT-Plus, especially in immunosuppressive patients where sensitivity was 100%. Further studies are warranted to confirm these results in broader populations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Latent TuberculosisInterferon-gamma Release TestsMycobacterium tuberculosisFluorescent Antibody TechniqueHumansMaleFemaleAdultMiddle AgedSensitivity and SpecificityYoung AdultReproducibility of ResultsTuberculin TestInterferon-gammaAged

Substanzen

Interferon-gamma

Zitieren

APA

Ceylan, A. N., Canat, H. D., Komec, S., & Gulmez, A. (2025). Diagnostic Accuracy of a Novel Fluorescent Immunoassay-Based Test in Latent Tuberculosis Infection. Clinical Laboratory. https://doi.org/10.7754/clin.lab.2024.241003

Vancouver

Ceylan AN, Canat HD, Komec S, Gulmez A. Diagnostic Accuracy of a Novel Fluorescent Immunoassay-Based Test in Latent Tuberculosis Infection. Clinical Laboratory. 2025. doi:10.7754/clin.lab.2024.241003

BibTeX
@article{Ceylan2025,
  author = {Ceylan, Ayse N and Canat, Harbiye D and Komec, Selda and Gulmez, Abdurrahman},
  title = {Diagnostic Accuracy of a Novel Fluorescent Immunoassay-Based Test in Latent Tuberculosis Infection},
  journal = {Clinical Laboratory},
  year = {2025},
  doi = {10.7754/clin.lab.2024.241003}
}