A Rare Case of Falsely Elevated High-Sensitivity Cardiac Troponin T due to Antibody Interference and a Literature Review
Clinical Laboratory · 2023
DOI: 10.7754/clin.lab.2023.230230 ↗Kurz gefasst
Übersichtsarbeit in der Zeitschrift Clinical Laboratory, 2023. Verfasst von Fu Cheng und 4 weiteren Autor:innen. Hauptthemen laut MEDLINE: Troponin T, Antibodies. Die Arbeit wurde bisher 2-mal zitiert (Crossref).
Abstract (Originalsprache)
BACKGROUND: This study aimed to demonstrate potential measures to discern interference and establish an algorithm for identifying interference in an hs-cTnT assay with a clinical discrepancy. METHODS: Laboratory error and disease factors were first excluded. Then, a serial dilution test, polyethylene glycol (PEG) precipitation, and heterophile antibody blocking reagents (HBRs) were performed to confirm the interference substance. In addition, high-speed centrifugation and western blot were tried. RESULTS: The sample diluted linearly (R2 = 0.9966). However, there was a dramatic reduction in the concentration of hs-cTnT concentration to a normal level after both PEG precipitation and HBR1 treatment. The results were inconclusive in high-speed centrifugation and western blot assay. CONCLUSIONS: This study elucidated a heterophile antibody interference in measurement of hs-cTnT alone on a Roche analyzer.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Zitieren
APA
Cheng, F., Chen, J., Wu, B. L., Yang, X., & Huang, Y. S. (2023). A Rare Case of Falsely Elevated High-Sensitivity Cardiac Troponin T due to Antibody Interference and a Literature Review. Clinical Laboratory. https://doi.org/10.7754/clin.lab.2023.230230
Vancouver
Cheng F, Chen J, Wu BL, Yang X, Huang YS. A Rare Case of Falsely Elevated High-Sensitivity Cardiac Troponin T due to Antibody Interference and a Literature Review. Clinical Laboratory. 2023. doi:10.7754/clin.lab.2023.230230
BibTeX
@article{Cheng2023,
author = {Cheng, Fu and Chen, Jie and Wu, Bing L and Yang, Xing and Huang, Yuan S},
title = {A Rare Case of Falsely Elevated High-Sensitivity Cardiac Troponin T due to Antibody Interference and a Literature Review},
journal = {Clinical Laboratory},
year = {2023},
doi = {10.7754/clin.lab.2023.230230}
}
EUVIMED