EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 3.0

Design of Extrusion Die for Medical Multi-lumen Tube Using Inverse Extrusion Simulation and Optimization

Yerim Kim, Kyungwook Ko, Wonjin Jun, Woojin Kim, Euntaek Lee

Journal of the Korean Society for Precision Engineering · 01.03.2026 · 43 (3) · 297-305

DOI: 10.7736/jkspe.025.120 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of the Korean Society for Precision Engineering, erschienen am 01.03.2026. Verfasst von Yerim Kim und 4 weiteren Autor:innen. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 3.0). Gefördert von: Ministry of Trade, Industry and Resources, Ministry of Education, Gyeongbuk RISE Center.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The design of the extrusion die significantly affects both the extrusion process and the quality of multi-lumen tubes. Traditional design methods that rely on trial and error tend to increase manufacturing time and costs while diminishing product quality. This study utilizes inverse extrusion simulation and optimization to design the extrusion die without the need for trial and error. The inverse extrusion simulation generates the die profile necessary to achieve the desired extrudate shape. Subsequently, direct extrusion simulations are conducted to predict the extrudate profile based on the derived die. The optimal volumetric flow rates of air within the lumens are also identified to ensure the extrudate meets the target profile. The results from the direct extrusion simulation, combined with optimization, confirm that the designed extrusion die can successfully produce the target profile. Using the derived die, the multi-lumen tube with the desired specifications is successfully extruded. This design and manufacturing approach enhances both the quality and productivity of multi-lumen tubes.

Förderung

  • Ministry of Trade, Industry and Resources
  • Ministry of Education
  • Gyeongbuk RISE Center

Zitieren

APA

Kim, Y., Ko, K., Jun, W., Kim, W., & Lee, E. (2026). Design of Extrusion Die for Medical Multi-lumen Tube Using Inverse Extrusion Simulation and Optimization. Journal of the Korean Society for Precision Engineering, 43(3), 297-305. https://doi.org/10.7736/jkspe.025.120

Vancouver

Kim Y, Ko K, Jun W, Kim W, Lee E. Design of Extrusion Die for Medical Multi-lumen Tube Using Inverse Extrusion Simulation and Optimization. Journal of the Korean Society for Precision Engineering. 2026;43(3):297-305. doi:10.7736/jkspe.025.120

BibTeX
@article{Kim2026,
  author = {Kim, Yerim and Ko, Kyungwook and Jun, Wonjin and Kim, Woojin and Lee, Euntaek},
  title = {Design of Extrusion Die for Medical Multi-lumen Tube Using Inverse Extrusion Simulation and Optimization},
  journal = {Journal of the Korean Society for Precision Engineering},
  year = {2026},
  volume = {43},
  number = {3},
  pages = {297--305},
  doi = {10.7736/jkspe.025.120}
}
Kostenlos starten →