Quantitative outperforms visual assessment of nonparallel orientation in ultrasound breast imaging reporting and data system
PeerJ · 18.03.2026 · 14 · e20992
DOI: 10.7717/peerj.20992 ↗Kurz gefasst
Fachartikel in der Zeitschrift PeerJ, online erschienen am 18.03.2026. Verfasst von Kailiang Chen und 2 weiteren Autor:innen. Beteiligte Einrichtung: The First Affiliated Hospital of Hainan Medical University, Haikou, Hainan Province, China. Hauptthemen laut MEDLINE: Breast Neoplasms, Ultrasonography, Mammary. Die Arbeit nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objective
The visual assessment of the “nonparallel orientation” descriptor in the ultrasound Breast Imaging Reporting and Data System (BI-RADS) is subjective and may affect both the accurate interpretation of breast masses and the overall diagnostic performance of the ultrasound BI-RADS. The objective of this study was to determine whether quantitative measurement of the nonparallel orientation descriptor improves diagnostic performance in the evaluation of breast malignancy assessment.
Methods
This prospective study, conducted at a tertiary hospital, analyzed 253 out of 6,893 patients with ultrasound BI-RADS 3-5 solid breast masses. For each mass, parallel or nonparallel orientation was assessed visually, and the orientation angle of breast mass was measured quantitatively on ultrasound image using built-in ultrasound software. Histopathological diagnosis served as the reference standard. Receiver Operating Characteristic (ROC) curve was plotted to determine the optimal cutoff value for the orientation angle in assessing breast malignancy. The diagnostic performances of the standard BI-RADS (using visual nonparallel orientation) and a modified BI-RADS (using the quantitative orientation angle) were compared for malignancy stratification.
Results
McNemar testing demonstrated significant differences in diagnostic outcomes between visual nonparallel orientation assessment and quantitative angle measurement, as well as between the standard and modified ultrasound BI-RADS classifications (all p < 0.001). The area under the curve (AUC) for visual assessment was 0.651, compared to 0.838 for the orientation angle ( p < 0.001). Incorporating orientation angle into ultrasound BI-RADS showed higher diagnostic performance (AUC = 0.922) compared to the standard ultrasound BI-RADS in this cohort (AUC = 0.905, p = 0.024).
Conclusion
The quantitative orientation angle is a more reproducible and objective measure. It can serve as a valuable complementary descriptor within ultrasound BI-RADS. Integrating this quantitative angle into a multi-descriptor assessment of breast masses improves the overall diagnostic performance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- The First Affiliated Hospital of Hainan Medical University, Haikou, Hainan Province, China.
- UWE College of Hainan Medical University, Haikou, Hainan Province, China.
Zitieren
APA
Chen, K., Chen, Q., & Wu, S. (2026). Quantitative outperforms visual assessment of nonparallel orientation in ultrasound breast imaging reporting and data system. PeerJ, 14, e20992. https://doi.org/10.7717/peerj.20992
Vancouver
Chen K, Chen Q, Wu S. Quantitative outperforms visual assessment of nonparallel orientation in ultrasound breast imaging reporting and data system. PeerJ. 2026;14:e20992. doi:10.7717/peerj.20992
BibTeX
@article{Chen2026,
author = {Chen, Kailiang and Chen, Qingfang and Wu, Size},
title = {Quantitative outperforms visual assessment of nonparallel orientation in ultrasound breast imaging reporting and data system},
journal = {PeerJ},
year = {2026},
volume = {14},
pages = {e20992},
doi = {10.7717/peerj.20992}
}
EUVIMED