Reliability and responsiveness of a tissue hardness meter and algometer for measuring tissue hardness and pressure pain threshold in upper trapezius myofascial trigger points

Soukmisai Somphithak, Uraiwan Chatchawan, Atipong Pimdee, Wiraphong Sucharit

PeerJ · 09.06.2025 · 13 · e19580

DOI: 10.7717/peerj.19580 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PeerJ, online erschienen am 09.06.2025. Verfasst von Soukmisai Somphithak und 3 weiteren Autor:innen. Beteiligte Einrichtung: School of Physical Therapy, Faculty of Associated Medical Sciences (AMS), Khon Kaen University (KKU), Khon Kaen, Thailand. Hauptthemen laut MEDLINE: Trigger Points, Superficial Back Muscles, Pain Threshold, Myofascial Pain Syndromes. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Research Center in Back, Neck, Other Joint Pain and Human Performance, Khon Kaen University (KKU), Khon Kaen, Thailand, Faculty of Associated Medical Sciences.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Tissue hardness meter and algometer (THA) are used to assess tissue hardness (TH) and pressure pain threshold (PPT), particularly in the evaluation of myofascial trigger points (MTrPs). This study introduces a side-lying protocol designed to comprehensively measure all portions of the upper trapezius (UT) muscle.

Purpose

The objective was to determine the reliability and responsiveness of THA to measure TH and PPT in patients with MTrPs in the UT muscle.

Methods

Reliability of TH and PPT measurements was assessed in a sample of 24 participants. Intra-rater and inter-rater reliability were evaluated using the intra-class correlation coefficient (ICC 3,1 ), while absolute reliability was established via Bland–Altman analysis, including the calculation of 95% limits of agreement (95% LoA). To assess responsiveness, 36 additional participants were recruited. Both distribution-based methods (mean difference, effect size (ES), standardized response mean (SRM), standard error of measurement (SEM), and minimal detectable change at 95% confidence (MDC 95 )) and anchor-based methods (minimal clinically important difference (MCID) and area under the curve (AUC)) were utilized in the analysis.

Results

Intra-rater reliability was excellent for both TH and PPT (ICC 3,1 : 0.95–0.97), while inter-rater reliability was moderate (ICC 3,1 : 0.60). Evidence of both fixed and proportional bias was identified for both TH and PPT. For TH, the SEM and MDC 95 were 2.66% and 7.37%, respectively, while for PPT, they were 0.12 kg/cm 2 and 0.34 kg/cm 2 , respectively. Following six physical therapy sessions, significant reductions in TH (mean: −7.86%; MCID: −7.43%; AUC: 0.97) and significant increases in PPT (mean: 0.20 kg/cm 2 ; MCID: 0.21 kg/cm 2 ; AUC: 0.86) were observed. Additionally, changes in PPT showed a negative correlation with improvements in the Neck Disability Index (NDI) (r = −0.35, p < 0.05).

Conclusion

The side-lying protocol demonstrated reliable and clinically relevant TH and PPT measurements, supporting its use for monitoring treatment outcomes in patients with MTrPs in the UT muscle.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Trigger PointsSuperficial Back MusclesPain ThresholdMyofascial Pain SyndromesPain MeasurementHumansMaleFemaleReproducibility of ResultsAdultMiddle AgedHardnessYoung AdultPressure

Stichwörter der Autor:innen

Myofascial trigger pointsPressure pain thresholdReliabilityResponsivenessSide-lying protocolTissue hardnessTissue hardness meter and algometerUpper trapezius muscle

Autor:innen und Institutionen

  • School of Physical Therapy, Faculty of Associated Medical Sciences (AMS), Khon Kaen University (KKU), Khon Kaen, Thailand.
  • Research Center in Back, Neck, Other Joint Pain and Human Performance (BNOJPH), Khon Kaen University (KKU), Khon Kaen, Thailand.

Förderung

  • Research Center in Back, Neck, Other Joint Pain and Human Performance
  • Khon Kaen University (KKU), Khon Kaen, Thailand
  • Faculty of Associated Medical Sciences

Zitieren

APA

Somphithak, S., Chatchawan, U., Pimdee, A., & Sucharit, W. (2025). Reliability and responsiveness of a tissue hardness meter and algometer for measuring tissue hardness and pressure pain threshold in upper trapezius myofascial trigger points. PeerJ, 13, e19580. https://doi.org/10.7717/peerj.19580

Vancouver

Somphithak S, Chatchawan U, Pimdee A, Sucharit W. Reliability and responsiveness of a tissue hardness meter and algometer for measuring tissue hardness and pressure pain threshold in upper trapezius myofascial trigger points. PeerJ. 2025;13:e19580. doi:10.7717/peerj.19580

BibTeX
@article{Somphithak2025,
  author = {Somphithak, Soukmisai and Chatchawan, Uraiwan and Pimdee, Atipong and Sucharit, Wiraphong},
  title = {Reliability and responsiveness of a tissue hardness meter and algometer for measuring tissue hardness and pressure pain threshold in upper trapezius myofascial trigger points},
  journal = {PeerJ},
  year = {2025},
  volume = {13},
  pages = {e19580},
  doi = {10.7717/peerj.19580}
}
Kostenlos starten →