EUVIMEDEUVIMED entdecken
Crossref

Information Disclosure, Medical Device Regulation, and Device Safety: The Case of Cook Celect IVC Filters

Kushal T Kadakia, Behnood Bikdeli, Aakriti Gupta, Sanket S Dhruva iD, Joseph S Ross iD, Harlan M Krumholz iD

Annals of Internal Medicine · 2024

DOI: 10.7326/annals-24-00089 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Annals of Internal Medicine, 2024. Verfasst von Kushal T Kadakia und 5 weiteren Autor:innen. Beteiligte Einrichtung: Harvard Medical School, Boston, Massachusetts (K.T.K.). Hauptthemen laut MEDLINE: Pulmonary Embolism, United States Food and Drug Administration, Vena Cava Filters. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 27 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Although medical devices are widely used in clinical practice, clinicians and the public have limited access to information on how devices are tested, regulated, and used, posing challenges to patient safety. This article uses Cook Medical's Celect inferior vena cava (IVC) filter, a medical device used for prevention of pulmonary embolism, as a case study of the transparency gap in medical device regulation. Recently unsealed court documents from litigation related to Celect reveal that the device's clinical study protocol did not follow U.S. Food and Drug Administration (FDA) guidance for IVC filter testing and that study outcome definitions for IVC perforation had lower sensitivity for detecting adverse events than those recommended by professional societies. Furthermore, a comparison of court documents and the public record indicates that adverse events and patient deaths were misreported to FDA reviewers and were inaccurately reported in the published literature and on the device label, providing patients and clinicians with inaccurate information about the device's safety. The Celect IVC filter case demonstrates the need for regulatory reforms to ensure that critical safety data are accessible to the FDA, clinicians, and patients to inform decision making.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pulmonary EmbolismUnited States Food and Drug AdministrationVena Cava FiltersHumansDevice ApprovalDisclosureEquipment SafetyPatient SafetyUnited States

Autor:innen und Institutionen

  • Harvard Medical School, Boston, Massachusetts (K.T.K.).
  • Center for Outcomes Research and Evaluation, Yale New Haven Hospital, New Haven, Connecticut; Cardiovascular Medicine Division, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts; and Thrombosis…
  • Smidt Heart Institute, Cedars-Sinai Medical Center, Los Angeles, California (A.G.).
  • Philip R. Lee Institute for Health Policy Studies, University of California, San Francisco, and Section of Cardiology, Department of Medicine, San Francisco Veterans Affairs Medical Center, San Francisco, California (S.…
  • Center for Outcomes Research and Evaluation, Yale New Haven Hospital; Section of General Internal Medicine and the National Clinician Scholars Program, Department of Internal Medicine, Yale School of Medicine; and Depar…
  • Center for Outcomes Research and Evaluation, Yale New Haven Hospital; Department of Health Policy and Management, Yale School of Public Health; and Section of Cardiovascular Medicine, Department of Internal Medicine, Ya…

ORCID-Profile: Sanket S Dhruva, Joseph S Ross, Harlan M Krumholz

Zitieren

APA

Kadakia, K. T., Bikdeli, B., Gupta, A., Dhruva, S. S., Ross, J. S., & Krumholz, H. M. (2024). Information Disclosure, Medical Device Regulation, and Device Safety: The Case of Cook Celect IVC Filters. Annals of Internal Medicine. https://doi.org/10.7326/annals-24-00089

Vancouver

Kadakia KT, Bikdeli B, Gupta A, Dhruva SS, Ross JS, Krumholz HM. Information Disclosure, Medical Device Regulation, and Device Safety: The Case of Cook Celect IVC Filters. Annals of Internal Medicine. 2024. doi:10.7326/annals-24-00089

BibTeX
@article{Kadakia2024,
  author = {Kadakia, Kushal T and Bikdeli, Behnood and Gupta, Aakriti and Dhruva, Sanket S and Ross, Joseph S and Krumholz, Harlan M},
  title = {Information Disclosure, Medical Device Regulation, and Device Safety: The Case of Cook Celect IVC Filters},
  journal = {Annals of Internal Medicine},
  year = {2024},
  doi = {10.7326/annals-24-00089}
}
Kostenlos starten →