Evaluation of the accuracy of a custom-made plate in maxillary repositioning
Archives of Craniofacial Surgery · 20.06.2026 · 27 (3) · 120-128
DOI: 10.7181/acfs.2025.0040 ↗Kurz gefasst
Fachartikel in der Zeitschrift Archives of Craniofacial Surgery, erschienen am 20.06.2026. Verfasst von Wafaa A Mohamed und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oral and Maxillofacial Surgery, Faculty of Dental Medicine for Girls, Al-Azhar University, Cairo, Egypt. wafaaanwar89@yahoo.com. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Background
Orthognathic surgery has become increasingly accurate with the integration of computerized technologies. This study aimed to evaluate the accuracy of custom-made plates in transferring the three-dimensional virtual plan to intraoperative maxillary repositioning.
Methods
This study included 10 patients. A digital surgical simulation technique was used for preoperative planning. The surgical components—osteotomy/plates, locating guides, and customized titanium plates—were fabricated using computer-aided design and manufacturing techniques, replacing the need for a surgical splint. Postoperative outcomes were assessed using unaltered anatomical landmarks from the preoperative virtual plan. Absolute discrepancies in translational and rotational movements between the planned and actual postoperative positions were measured through superimposition analysis. This evaluation assessed the precision of custom-made plates in transferring the virtual surgical plan, identifying deviations in the cephalocaudal, mediolateral, and anteroposterior directions that could influence functional and esthetic outcomes.
Results
No statistically significant differences were observed for any landmarks across the axial, coronal, and sagittal planes. The greatest variability was noted at the posterior nasal spine landmark in the axial plane (1.7 mm), compared with the coronal (0.9 mm) and sagittal (0.5 mm) planes.
Conclusion
Customized titanium plates effectively transferred the virtual surgical plan for maxillary repositioning without the need for an occlusal splint. However, potential posterior interference should be considered during surgical execution.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Oral and Maxillofacial Surgery, Faculty of Dental Medicine for Girls, Al-Azhar University, Cairo, Egypt. wafaaanwar89@yahoo.com.
- Department of Oral and Maxillofacial Surgery, Faculty of Dental Medicine for Girls, Al-Azhar University, Cairo, Egypt.
- Department of Oral and Maxillofacial Surgery, Deraya University, New Minia, Egypt.
- Department of Oral and Maxillofacial Surgery, Faculty of Dental Medicine, Future University, New Cairo, Egypt.
- Sharq El Madina Hospital, Alexandria, Egypt.
- Department of Orthodontics, Faculty of Dental Medicine, Menoufia University, Shebin El-Kom, Egypt.
Zitieren
APA
Mohamed, W. A., Hassan, S. A., Mohamed, F. I., Barakat, A. A., Alahmady, H. H., Fata, M. M., & Mohamed, A. A. (2026). Evaluation of the accuracy of a custom-made plate in maxillary repositioning. Archives of Craniofacial Surgery, 27(3), 120-128. https://doi.org/10.7181/acfs.2025.0040
Vancouver
Mohamed WA, Hassan SA, Mohamed FI, Barakat AA, Alahmady HH, Fata MM, et al. Evaluation of the accuracy of a custom-made plate in maxillary repositioning. Archives of Craniofacial Surgery. 2026;27(3):120-128. doi:10.7181/acfs.2025.0040
BibTeX
@article{Mohamed2026,
author = {Mohamed, Wafaa A and Hassan, Susan A and Mohamed, Fatma I and Barakat, Ahmed A and Alahmady, Hatem H and Fata, Mahmoud M and Mohamed, Ahmed A},
title = {Evaluation of the accuracy of a custom-made plate in maxillary repositioning},
journal = {Archives of Craniofacial Surgery},
year = {2026},
volume = {27},
number = {3},
pages = {120--128},
doi = {10.7181/acfs.2025.0040}
}
EUVIMED