EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

Diseño de lentes intraoculares tóricas y evaluación de su calidad óptica en ojo modelo

Aser Cebollada-Martínez, Kevin Moldvai, Laura Remon Martin, Carla Sierra Santamaria, Marta Sancho Larraz, Blanca Fernandez Gonzalez, Patricia Ponce Cabrera

Optometría Clínica y Ciencias de la Visión · 2024

DOI: 10.71413/7ahpbp72 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Optometría Clínica y Ciencias de la Visión, 2024. Verfasst von Aser Cebollada-Martínez und 6 weiteren Autor:innen. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Relevance: The study evaluates the image quality of toric intraocular lenses (IOLs-T) with different levels of toricity and aspheric geometries, comparing them with spherical IOLs-T. It analyzes their tolerance to tilt, decentration, and rotation in a model eye through numerical simulations, without experimental testing.

Purpose

To assess the image quality of various designs of toric intraocular lenses (IOLs-T) and their tolerance to tilt, decentration, and rotation in a model eye. IOLs-T with different degrees of toricity and aspheric geometries (negative and neutral asphericities) were designed, adjusting the model's corneal toricity to compensate for the IOLs' toricity. The results, obtained numerically without experimental testing, compare the optical performance of IOLs with varying toricities and asphericities, including spherical toric IOLs.

Methods

To evaluate the optical performance of toric intraocular lenses (IOLs-T) under different conditions, a numerical model of a pseudophakic eye was used in ZEMAX® software. Based on Atchinson's schematic eye, the study considered IOLs-T with various astigmatic powers and aspheric geometries (neutral and negative). The effects of decentration, tilt, and rotation of the IOLs-T on image quality were analyzed by measuring Modulation Transfer Function (MTF) and Zernike coefficients. The results suggest that the retina's position affects how these factors impact optical quality.

Results

The study evaluated the optical performance of various designs of toric intraocular lenses (IOLs-T) under conditions of decentration, tilt, and rotation, using metrics such as Modulation Transfer Function (MTF) and root mean square (RMS) error of aberrations. Lens B showed the best optical quality when centered but was most affected by decentration and rotation, with a significant drop in MTF. Lenses A and spherical IOLs initially offered lower quality but greater stability against displacements. Lens B had minimal defocus when centered and performed better with tilts up to 4°, whereas other lenses experienced greater MTF reduction with smaller tilts. These results indicate that while Lens B provides the best centered optical quality, its performance significantly deteriorates with decentration and tilt.

Conclusions

Toric IOLs with higher conicity constants offer better optical performance in terms of MTF and wavefront aberration when centered, surpassing spherical lenses. Although MTF is independent of cylinder power, Lens B is more sensitive to decentration, while lenses with lower conicity or spherical lenses are less affected. Astigmatism and coma aberrations are similar among designs under tilt or rotation. The results suggest that IOLs with minimal aberrations may be the best option if perfect alignment cannot be ensured.

Zitieren

APA

Cebollada-Martínez, A., Moldvai, K., Martin, L. R., Santamaria, C. S., Larraz, M. S., Gonzalez, B. F., & Cabrera, P. P. (2024). Diseño de lentes intraoculares tóricas y evaluación de su calidad óptica en ojo modelo. Optometría Clínica y Ciencias de la Visión. https://doi.org/10.71413/7ahpbp72

Vancouver

Cebollada-Martínez A, Moldvai K, Martin LR, Santamaria CS, Larraz MS, Gonzalez BF, et al. Diseño de lentes intraoculares tóricas y evaluación de su calidad óptica en ojo modelo. Optometría Clínica y Ciencias de la Visión. 2024. doi:10.71413/7ahpbp72

BibTeX
@article{CebolladaMartnez2024,
  author = {Cebollada-Martínez, Aser and Moldvai, Kevin and Martin, Laura Remon and Santamaria, Carla Sierra and Larraz, Marta Sancho and Gonzalez, Blanca Fernandez and Cabrera, Patricia Ponce},
  title = {Diseño de lentes intraoculares tóricas y evaluación de su calidad óptica en ojo modelo},
  journal = {Optometría Clínica y Ciencias de la Visión},
  year = {2024},
  doi = {10.71413/7ahpbp72}
}
Kostenlos starten →