Design and Development of Drybath Incubator for Chemical Test Tubes with Fuzzy Logic Controller
Electromedic, Journal of Medical Electronic · 04.06.2026 · 3 (1)
DOI: 10.66508/electromedic.v3i1.139 ↗Kurz gefasst
Fachartikel in der Zeitschrift Electromedic, Journal of Medical Electronic, online erschienen am 04.06.2026. Verfasst von Hendra Yoga und 2 weiteren Autor:innen.
Abstract (Originalsprache)
Clinical Pathology Laboratories require equipment for incubating samples and chemical reagents at temperatures between 25°C and 45°C. Commonly used waterbaths have limitations including contamination risk, temperature inaccuracy, and lack of incubation time display. Additionally, non-standard incubation practices such as holding test tubes in hands are often performed, reducing test result precision. This study aimed to design and build a drybath incubator for chemical test tubes with a fuzzy logic-based temperature control system and integrated timer. The system employs a hollow aluminum block, cartridge heater, waterproof DS18B20 temperature sensor, Arduino Mega microcontroller, and a 3.5-inch TFT LCD touchscreen. The fuzzy logic controller regulates temperature from 25°C to 45°C. Testing included accuracy, stability, temperature distribution uniformity, and timer precision tests across 25 holes (20 holes Ø 1.25 cm and 5 holes Ø 1.5 cm). Results showed temperature accuracy of 99.99%–100.00%, temperature stability with fluctuations below ±0.4°C, uniform temperature distribution with 99.98%–100.00% accuracy, and timer accuracy of 99.994%. The system proved capable of replacing waterbaths as incubators for chemical reactions in clinical pathology laboratories.
Zitieren
APA
Yoga, H., Nurdinawati, V., & Hendryani, A. (2026). Design and Development of Drybath Incubator for Chemical Test Tubes with Fuzzy Logic Controller. Electromedic, Journal of Medical Electronic, 3(1). https://doi.org/10.66508/electromedic.v3i1.139
Vancouver
Yoga H, Nurdinawati V, Hendryani A. Design and Development of Drybath Incubator for Chemical Test Tubes with Fuzzy Logic Controller. Electromedic, Journal of Medical Electronic. 2026;3(1). doi:10.66508/electromedic.v3i1.139
BibTeX
@article{Yoga2026,
author = {Yoga, Hendra and Nurdinawati, Vita and Hendryani, Atika},
title = {Design and Development of Drybath Incubator for Chemical Test Tubes with Fuzzy Logic Controller},
journal = {Electromedic, Journal of Medical Electronic},
year = {2026},
volume = {3},
number = {1},
doi = {10.66508/electromedic.v3i1.139}
}