Development of a PETG-based 3D-printed lumbar spine phantom for bone mineral density assessment
Journal of Associated Medical Sciences · 05.08.2026 · 59 (3) · 344-356
DOI: 10.66285/jams.2026.107 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Associated Medical Sciences, online erschienen am 05.08.2026. Verfasst von Thunyarat Chusin und 5 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background
Recent advances in three-dimensional printing technology have enabled the development of customized phantoms for quantitative imaging applications, including bone mineral density assessment.
Objectives
This study aimed to develop and validate a 3D-printed lumbar spine phantom by comparing its radiological characteristics with those of the commercial anthropomorphic RSD RS-113T pelvic phantom and evaluating its performance against the GE Lunar iDXA quality control (QC) phantom. Materials and methods: A commercial RSD RS-113T pelvic phantom was scanned using CT and DXA to obtain reference CT numbers and BMD values. Lumbar spine structures were segmented from CT images to design a 3D-printed lumbar spine phantom using fused deposition modeling (FDM). Polyethylene terephthalate glycol (PETG; densities 1.24 and 1.27 gm/cm³), combined with calcium carbonate powder, was used to simulate bone mineral content. CT number and BMD measurements of the developed phantoms were compared with those of the RSD RS-113T pelvic phantom using Bland-Altman agreement analysis. BMD repeatability of the developedphantoms was subsequently validated using the GE Lunar iDXA QC phantom.
Results
The PETG black phantom (1.27 gm/cm³) showed closer agreement with the RSD RS-113T pelvic phantom than the PETG white phantom (1.24 gm/cm³), with smaller Bland–Altman mean biases for CT number (8.65 vs -297.10 HU) and BMD (0.22 vs 0.84 gm/cm²). The RSD RS-113T pelvic phantom exhibited BMD CV values of less than 2%. The PETG black phantom also demonstrated better BMD repeatability (CV: 0.58%-2.92%) than the PETG white phantom, although some measurements exceeded the 2% precision criterion.
Conclusion
The PETG black 3D-printed lumbar spine phantom demonstrated superior BMD repeatability compared with the PETG white phantom. It also showed good agreement with the RSD RS-113T pelvic phantom and satisfactory BMD performance when validated using the GE Lunar iDXA QC phantom.
Zitieren
APA
Chusin, T., Kongfun, P., Chanapim, S., Thanashappokin, S., Pengpan, T., & Aphiwatthanasumet, K. (2026). Development of a PETG-based 3D-printed lumbar spine phantom for bone mineral density assessment. Journal of Associated Medical Sciences, 59(3), 344-356. https://doi.org/10.66285/jams.2026.107
Vancouver
Chusin T, Kongfun P, Chanapim S, Thanashappokin S, Pengpan T, Aphiwatthanasumet K. Development of a PETG-based 3D-printed lumbar spine phantom for bone mineral density assessment. Journal of Associated Medical Sciences. 2026;59(3):344-356. doi:10.66285/jams.2026.107
BibTeX
@article{Chusin2026,
author = {Chusin, Thunyarat and Kongfun, Pakaphorn and Chanapim, Suchanya and Thanashappokin, Suphichaya and Pengpan, Thanyawee and Aphiwatthanasumet, Kingkarn},
title = {Development of a PETG-based 3D-printed lumbar spine phantom for bone mineral density assessment},
journal = {Journal of Associated Medical Sciences},
year = {2026},
volume = {59},
number = {3},
pages = {344--356},
doi = {10.66285/jams.2026.107}
}
EUVIMED