Method comparison and clinical classification agreement between Freelite and N-Latex serum-free light chain assays for clinical laboratory implementation

Dararat Dankai, Phichayut Phinyo, Wiyada Dankai

Journal of Associated Medical Sciences · 14.07.2026 · 59 (3) · 256-267

DOI: 10.66285/jams.2026.099 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Associated Medical Sciences, online erschienen am 14.07.2026. Verfasst von Dararat Dankai und 2 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Measurement of serum free light chains (SFLC) is an essential component in the diagnosis and monitoring of plasma cell dyscrasias. Two analytical platforms are widely used in clinical laboratories: the Freelite assay, which employs polyclonal antibodies with turbidimetric detection, and the N-Latex assay, which uses monoclonal antibodies with nephelometric measurement. Because these assays differ in antibody design, calibration, and analytical detection principles, numerical concentrations and κ/λ ratios maynot be directly interchangeable across platforms. This study aimed to evaluate the analytical agreement and clinical classification concordance between the Freelite and N-Latex assays to assess the feasibility of implementing the N-Latex method in routine laboratory practice. Materials and methods: A cross-sectional method-comparison study was performed using 33 paired archived residual serum samples obtained from routine clinical testing. Free κ and λ concentrations were measured using both Freelite and N-Latex assays on the same specimens. Analytical agreement was assessed using the Wilcoxon signed-rank test, Spearman correlation, Passing–Bablok regression, and Bland-Altman analysis. κ/λ ratios were interpreted according to assay-specific reference intervals, and categorical agreement was evaluated using Cohen’s kappa coefficient.

Results

Freelite produced significantly higher κ concentrations than N-Latex (median 43.0 vs 38.9 mg/L, p=0.0002), whereas λ concentrations did not differ significantly. Strong correlations were observed for both κ (ρ=0.897) and λ (ρ=0.951); however, correlation alone was not sufficient to demonstrate quantitative interchangeability. Passing-Bablok regression demonstrated proportional bias for κ measurements, with Freelite yielding higher values at elevated concentrations and the 95% confidence interval of the slope not including 1.0. Bland-Altman analysis revealed wide limits of agreement for κ values at extreme concentrations, indicating clinically relevant numerical differences between methods. When κ/λ ratios were interpreted categorically, using assay-specific reference intervals, concordant classification was observed in 30 of 33 samples (91%), corresponding to substantial agreement (κ=0.79).

Conclusions

The Freelite and N-Latex assays demonstrate strong correlation but exhibit systematic quantitative differences, particularly for κ measurements at high concentrations. Although κ/λ ratio interpretation is largely concordant, the two assays should not be considered directly interchangeable for absolute numerical interpretation or longitudinal monitoring. Implementation of the N-Latex assay is feasible in routine clinical laboratories provided that assay-specific reference intervals, appropriate dilution protocols, and antigen-excess precautions are applied, and clinicians are informed about expected differences when transitioning betweenanalytical platforms.

Zitieren

APA

Dankai, D., Phinyo, P., & Dankai, W. (2026). Method comparison and clinical classification agreement between Freelite and N-Latex serum-free light chain assays for clinical laboratory implementation. Journal of Associated Medical Sciences, 59(3), 256-267. https://doi.org/10.66285/jams.2026.099

Vancouver

Dankai D, Phinyo P, Dankai W. Method comparison and clinical classification agreement between Freelite and N-Latex serum-free light chain assays for clinical laboratory implementation. Journal of Associated Medical Sciences. 2026;59(3):256-267. doi:10.66285/jams.2026.099

BibTeX
@article{Dankai2026,
  author = {Dankai, Dararat and Phinyo, Phichayut and Dankai, Wiyada},
  title = {Method comparison and clinical classification agreement between Freelite and N-Latex serum-free light chain assays for clinical laboratory implementation},
  journal = {Journal of Associated Medical Sciences},
  year = {2026},
  volume = {59},
  number = {3},
  pages = {256--267},
  doi = {10.66285/jams.2026.099}
}
Kostenlos starten →