Continuous Hemodynamic Monitoring of Radial Artery Based on Flexible Piezoelectric Fabric Sensors
International Journal of Computational and Biological Sciences · 31.01.2026 · 3 (1) · 107-114
DOI: 10.66238/ijcbs49 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Computational and Biological Sciences, online erschienen am 31.01.2026. Verfasst von Ye-Eun Yoon und 3 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Continuous monitoring of hemodynamic parameters offers critical insights intocardiovascular health, potentially preventing acute events through early detection ofphysiological anomalies. Traditional cuff-based sphygmomanometers provide onlyintermittent readings, while invasive arterial catheterization, though accurate, carriessignificant risks and is unsuitable for ambulatory settings. This paper presents thedevelopment and characterization of a flexible, textile-based piezoelectric sensor designedfor the continuous acquisition of radial artery pulse waveforms. Utilizing polyvinylidenefluoride (PVDF) nanofibers integrated into abreathable fabric matrix, the sensor exhibitshigh electromechanical sensitivity and excellent conformity to the curvilinear surface of thehuman wrist. We detail the fabrication process involving electrospinning and ultrasonicbonding, followed by a comprehensive analysis of the sensor's electrical and mechanicalproperties. The device demonstrates the capability to capture high-fidelity arterial pressurewaves, including subtle features such as the dicrotic notch, which are essential forcalculating arterial stiffness and cardiac output. Signal processing algorithms weredeveloped to mitigate motion artifacts and baseline drift, ensuring robust performanceduring daily activities. Comparative studies against clinical-grade tonometers reveal a highdegree of correlation, suggesting that this fabric-based approach represents a viable noninvasive alternative for long-term hemodynamic monitoring.
Zitieren
APA
Yoon, Y., Takahashi, A., Davis, M., & Williams, S. (2026). Continuous Hemodynamic Monitoring of Radial Artery Based on Flexible Piezoelectric Fabric Sensors. International Journal of Computational and Biological Sciences, 3(1), 107-114. https://doi.org/10.66238/ijcbs49
Vancouver
Yoon Y, Takahashi A, Davis M, Williams S. Continuous Hemodynamic Monitoring of Radial Artery Based on Flexible Piezoelectric Fabric Sensors. International Journal of Computational and Biological Sciences. 2026;3(1):107-114. doi:10.66238/ijcbs49
BibTeX
@article{Yoon2026,
author = {Yoon, Ye-Eun and Takahashi, Akiko and Davis, Michael and Williams, Sarah},
title = {Continuous Hemodynamic Monitoring of Radial Artery Based on Flexible Piezoelectric Fabric Sensors},
journal = {International Journal of Computational and Biological Sciences},
year = {2026},
volume = {3},
number = {1},
pages = {107--114},
doi = {10.66238/ijcbs49}
}
EUVIMED