Lightweight Markerless AR Registration Using Depth Information for Endoscopic Procedure
International Journal of Computational and Biological Sciences · 01.12.2025 · 2 (02) · 160-165
DOI: 10.66238/ijcbs22 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Computational and Biological Sciences, online erschienen am 01.12.2025. Verfasst von Daniel R. McLean und 3 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Endoscopic surgeries require lightweight AR registration for constrained devices. We propose a sparse depth sampling algorithm combined with optimized ICP and pose graph pruning. Depth clouds are compressed to 15% of full resolution, reducing computational load. Tested on 200 in vivo endoscopic frames, mean error was 1.6 mm, with frame rates of 31 fps on mobile GPUs. Compared to dense ICP, runtime decreased by 62% with only 0.3 mm accuracy loss, enabling real-time deployment
Zitieren
APA
McLean, D. R., Kapoor, P. S., Carter, J. L., & Hughes, O. K. (2025). Lightweight Markerless AR Registration Using Depth Information for Endoscopic Procedure. International Journal of Computational and Biological Sciences, 2(02), 160-165. https://doi.org/10.66238/ijcbs22
Vancouver
McLean DR, Kapoor PS, Carter JL, Hughes OK. Lightweight Markerless AR Registration Using Depth Information for Endoscopic Procedure. International Journal of Computational and Biological Sciences. 2025;2(02):160-165. doi:10.66238/ijcbs22
BibTeX
@article{McLean2025,
author = {McLean, Daniel R. and Kapoor, Priya S. and Carter, Jonathan L. and Hughes, Olivia K.},
title = {Lightweight Markerless AR Registration Using Depth Information for Endoscopic Procedure},
journal = {International Journal of Computational and Biological Sciences},
year = {2025},
volume = {2},
number = {02},
pages = {160--165},
doi = {10.66238/ijcbs22}
}