Comparison of radiation exposure between pulsed and continuous fluoroscopy during PFN-A fixation of pertrochanteric fractures

Mehmet Ali Sabır, Rıza Berker Özbek, Recep Taşkin, Fatih Uğur, Ahmet Onur Çelik

Kastamonu Medical Journal · 16.06.2026 · 6 (2) · 136-141

DOI: 10.66235/kumj.1855797 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Kastamonu Medical Journal, online erschienen am 16.06.2026. Verfasst von Mehmet Ali Sabır und 4 weiteren Autor:innen. Die Arbeit nennt 21 Literaturangaben. Gefördert von: None..

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aims

Fluoroscopy is widely used in orthopedic surgery but raises concerns regarding radiation exposure. This study aimed to compare radiation exposure between pulsed and continuous fluoroscopy modes during PFN-A (proximal femoral nail - antirotation) fixation of pertrochanteric femoral fractures.

Methods

In this retrospective cohort study, 102 patients treated with PFN-A for AO/OTA (Arbeitsgemeinschaft für osteosynthesefragen Foundation/Orthopedic Trauma Association) type 31A1.2 and 31A1.3 pertrochanteric fractures at a single institution between January 2024 and January 2025 were included.

Results

No significant differences were observed between groups in terms of fluoroscopy time, total radiation dose, DAP (dose-area product), or number of exposures. However, radiation exposure per image was significantly lower in the pulsed fluoroscopy group, as demonstrated by both dose per image (p = 0.001, r = 0.32) and DAP per image (p = 0.002, r = 0.29). In adjusted analyses, fluoroscopy mode remained a significant predictor of radiation exposure per image (F = 9.07, p = 0.003). Image quality was considered subjectively adequate based on intraoperative assessment, with no clinically relevant difference between the groups. Radiation exposure was also normalized per image. Statistical analyses were performed using the Mann–Whitney U test and ANCOVA.

Conclusion

Pulsed fluoroscopy does not significantly reduce total radiation parameters but is associated with lower radiation exposure per image during PFN-A fixation of pertrochanteric fractures. These findings indicate a reduction in radiation exposure per image without significant differences in total radiation parameters.

Förderung

  • None.

Zitieren

APA

Sabır, M. A., Özbek, R. B., Taşkin, R., Uğur, F., & Çelik, A. O. (2026). Comparison of radiation exposure between pulsed and continuous fluoroscopy during PFN-A fixation of pertrochanteric fractures. Kastamonu Medical Journal, 6(2), 136-141. https://doi.org/10.66235/kumj.1855797

Vancouver

Sabır MA, Özbek RB, Taşkin R, Uğur F, Çelik AO. Comparison of radiation exposure between pulsed and continuous fluoroscopy during PFN-A fixation of pertrochanteric fractures. Kastamonu Medical Journal. 2026;6(2):136-141. doi:10.66235/kumj.1855797

BibTeX
@article{Sabr2026,
  author = {Sabır, Mehmet Ali and Özbek, Rıza Berker and Taşkin, Recep and Uğur, Fatih and Çelik, Ahmet Onur},
  title = {Comparison of radiation exposure between pulsed and continuous fluoroscopy during PFN-A fixation of pertrochanteric fractures},
  journal = {Kastamonu Medical Journal},
  year = {2026},
  volume = {6},
  number = {2},
  pages = {136--141},
  doi = {10.66235/kumj.1855797}
}
Kostenlos starten →