``Evaluation of Molybdenum Content in 99Mo/99mTc Generators Eluate''

Sunnia Shafiq, Hafiz Khush Naseeb Ahmad, Hina Manzoor, SadiqUllah SadiqUllah, Hashir Saeed, Muhammad Sadiq, Shehla Iftikhar

Journal of King Abdulaziz University: Engineering Sciences · 21.05.2026 · 36 (1) · 1-6

DOI: 10.64064/1658-4260.1018 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of King Abdulaziz University: Engineering Sciences, online erschienen am 21.05.2026. Verfasst von Sunnia Shafiq und 6 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

This study aimed to assess the reliability and purity of technetium-99m (99mTc) generators produced by the Pakistan Institute of Nuclear Science and Technology (PINSTECH), known as PAKGEN, by evaluating the molybdenum (99Mo) breakthrough levels over five years. Maintaining low 99Mo levels is crucial for patient safety and high-quality diagnostic imaging in nuclear medicine.

Methods

189 PAKGEN 99m Tc generators were tested at the Centre for Nuclear Medicine and Radiotherapy (CENAR) in Quetta, Pakistan, from December 2019 to December 2023. Each generator was eluated daily from Monday to Saturday, and the molybdenum content in the elution was measured using the dose calibrator Capintec CRC-55t. The results were compared to the limits set by the European and American Pharmacopeias, which allow 1 μ Ci and 0.15 μ Ci of 99Mo per mCi of 99m Tc, respectively. Statistical analysis was performed using Origin software.

Results

The molybdenum content in all tested generators remained well below the pharmacopeia limits throughout the study period. None of the 189 PAKGEN generators tested showed Mo breakthrough levels exceeding the American Pharmacopeia limit of 0.15 μ Ci/mCi or the European Pharmacopeia limit of 1.0 μ Ci/mCi. The consistently low levels of Mo contamination highlight the reliability of PAKGEN generators in maintaining high purity standards.

Conclusion

The study demonstrates that PAKGEN 99mTc generators produced by PINSTECH reliably maintain molybdenum breakthrough levels within safe limits, as the European and American Pharmacopeias define. This ensures high-quality diagnostic imaging and patient safety, underscoring the importance of stringent quality control measures in producing 99mTc generators. These findings are particularly significant for medical institutions in regions with limited access to advanced technologies, such as CENAR in Quetta, Pakistan.

Zitieren

APA

Shafiq, S., Ahmad, H. K. N., Manzoor, H., SadiqUllah, S., Saeed, H., Sadiq, M., & Iftikhar, S. (2026). ``Evaluation of Molybdenum Content in 99Mo/99mTc Generators Eluate''. Journal of King Abdulaziz University: Engineering Sciences, 36(1), 1-6. https://doi.org/10.64064/1658-4260.1018

Vancouver

Shafiq S, Ahmad HKN, Manzoor H, SadiqUllah S, Saeed H, Sadiq M, et al. ``Evaluation of Molybdenum Content in 99Mo/99mTc Generators Eluate''. Journal of King Abdulaziz University: Engineering Sciences. 2026;36(1):1-6. doi:10.64064/1658-4260.1018

BibTeX
@article{Shafiq2026,
  author = {Shafiq, Sunnia and Ahmad, Hafiz Khush Naseeb and Manzoor, Hina and SadiqUllah, SadiqUllah and Saeed, Hashir and Sadiq, Muhammad and Iftikhar, Shehla},
  title = {``Evaluation of Molybdenum Content in 99Mo/99mTc Generators Eluate''},
  journal = {Journal of King Abdulaziz University: Engineering Sciences},
  year = {2026},
  volume = {36},
  number = {1},
  pages = {1--6},
  doi = {10.64064/1658-4260.1018}
}
Kostenlos starten →