Background
Standard visual acuity tests prioritize central spatial resolution but consistently fail to identify functional peripheral visual deficits that impact spatial orientation and mobility. This cross-sectional observational study introduces and validates a novel Cubic Neuro-Visual Functional Screening Device designed to assess peripheral visual perception and central-peripheral visual integration dynamically.
Methods
One hundred and fifty participants aged 5–60 years were enrolled across five clinical cohorts: healthy controls (n=50), amblyopia (n=30), glaucoma (n=35), stroke-related peripheral field defects (n=20), and peripheral perception disturbance with normal visual acuity (n=15). The primary outcome was peripheral stimulus recognition response time, measured via a digital precision chronometer while the patient maintained central fixation, monitored by a micro-camera.
Results
Healthy controls achieved a mean response time of 0.78 ± 0.12 s, demonstrating significantly faster processing compared to an aggregate pathological group mean of 1.38 ± 0.25 s (p < 0.05). A crucial clinical finding emerged in patients with normal central visual acuity (6/6) but functional disturbances, who exhibited significant peripheral processing delays (1.20 ± 0.15 s). Furthermore, distinct behavioral compensatory mechanisms, including micro-saccades and head tilting, were systematically observed in pathological cohorts. Gender did not significantly influence response times.
Conclusions
The cubic screening device demonstrates substantial clinical potential in differentiating healthy from pathological peripheral visual processing. The findings highlight a critical diagnostic gap in relying solely on Snellen acuity, emphasizing the necessity for accessible, rapid functional peripheral assessment tools in routine clinical and neuro-visual rehabilitation settings.
Zitieren
APA
Al-Ani, M. S. A., Abdullah, S. A., Mohammed, M. H., & Maeedi, A. H. (2026). Evaluation of Peripheral Visual Perception Using a Cubic Neuro-Visual Functional Screening Device. Journal of Progressive Medical Sciences, 2(5). https://doi.org/10.63939/7ky4r309
Vancouver
Al-Ani MSA, Abdullah SA, Mohammed MH, Maeedi AH. Evaluation of Peripheral Visual Perception Using a Cubic Neuro-Visual Functional Screening Device. Journal of Progressive Medical Sciences. 2026;2(5). doi:10.63939/7ky4r309
BibTeX
@article{AlAni2026,
author = {Al-Ani, Mohaimen Samir Arif and Abdullah, Sajad Abbas and Mohammed, Marrwan Hisham and Maeedi, Ali Hadi},
title = {Evaluation of Peripheral Visual Perception Using a Cubic Neuro-Visual Functional Screening Device},
journal = {Journal of Progressive Medical Sciences},
year = {2026},
volume = {2},
number = {5},
doi = {10.63939/7ky4r309}
}
EUVIMED