EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Optimizing Internal Target Margins for Moving Lung Tumors in VMAT-SBRT: A Phantom Study on Anatomical Modulation Complexity

Ahmet Fatih YÜREKLİ, Bahar DİRİCAN

Advances in Medical Pyhsics and Applied Sciences · 2026

DOI: 10.63187/ampas.58 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Advances in Medical Pyhsics and Applied Sciences, 2026. Verfasst von Ahmet Fatih YÜREKLİ und 1 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose: Target motion in VMAT-based lung SBRT plans complicates dosimetric quality assurance (QA). The aim of this study is to investigate the effect of tumor locations in different anatomical lung lobes and changes in the planning target volume (PTV) on static (breath-hold) and dynamic (free-breathing) QA results. Methodology: In this study, Monte Carlo (MC)-based VMAT SBRT plans were generated for five different lung lobes (Left Lower [LL], Left Upper [LU], Right Lower [RL], Right Middle [RM], Right Upper [RU]) using an anthropomorphic CIRS phantom. Measurements were performed both statically and dynamically (at respiratory frequencies of 0.3 Hz and 0.4 Hz and at various amplitudes) using a 2D array detector (Matrixx, 1%/1 mm) mounted on the custom-built respiratory motion simulator, as well as using an EPID (Dosisoft, 3%/1 mm). The data were analyzed using IBM SPSS Statistics v27.0.1.0 software with the Friedman and Kruskal-Wallis tests. Findings and Conclusion: The dosimetric distortion caused by tumor motion and margin expansion depends on the target’s anatomical proximity and the complexity of MLC modulation. Our study recommends the use of a two-stage “Hybrid QA” protocol, which combines EPID flux verification and the dynamic Matrixx motion test, to distinguish between planning (dosimetric) errors and mechanical synchronization (interplay) errors.

Zitieren

APA

YÜREKLİ, A. F., & DİRİCAN, B. (2026). Optimizing Internal Target Margins for Moving Lung Tumors in VMAT-SBRT: A Phantom Study on Anatomical Modulation Complexity. Advances in Medical Pyhsics and Applied Sciences. https://doi.org/10.63187/ampas.58

Vancouver

YÜREKLİ AF, DİRİCAN B. Optimizing Internal Target Margins for Moving Lung Tumors in VMAT-SBRT: A Phantom Study on Anatomical Modulation Complexity. Advances in Medical Pyhsics and Applied Sciences. 2026. doi:10.63187/ampas.58

BibTeX
@article{YREKL2026,
  author = {YÜREKLİ, Ahmet Fatih and DİRİCAN, Bahar},
  title = {Optimizing Internal Target Margins for Moving Lung Tumors in VMAT-SBRT: A Phantom Study on Anatomical Modulation Complexity},
  journal = {Advances in Medical Pyhsics and Applied Sciences},
  year = {2026},
  doi = {10.63187/ampas.58}
}
Kostenlos starten →