REAL-WORLD SAFETY AND PERFORMANCE OF THE COLOR DOPPLER ULTRASOUND SYSTEM: A MULTI-CENTER POST-MARKET CLINICAL FOLLOW-UP SURVEY

YuanBo Wang, FuCheng Wang, Dan Wu, JieBing Ma, BaoLi Zhang, Wen Xie

Journal of Pharmaceutical and Medical Research · 2026

DOI: 10.61784/jpmr3093 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Pharmaceutical and Medical Research, 2026. Verfasst von YuanBo Wang und 5 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To evaluate the real-world clinical safety and performance of the Color Doppler Ultrasound System (Model: M30/ViV 20, Manufacturer: Zoncare, China) across multiple centers and diverse operator profiles.

Methods

This cross-sectional PMCF survey enrolled evaluators from 61 centers. A standardized questionnaire was used to assess device safety and performance across four domains: Safety (2 items), Image Quality (6 items), Operational Experience (4 items), and Stability (3 items). The primary endpoint was the acceptability rate (dichotomized as "Acceptable" vs. "Not acceptable") for each indicator, with exact 95% confidence intervals (CIs) calculated. The secondary endpoint comprised the distribution of original ordinal responses (Likert-scale) and median scores for each item.

Results

A total of 76 valid questionnaires were returned. For the primary endpoint, the safety incidence (electric shock/skin burns) was 0% (95% CI: 0.0% - 4.7%). The overall acceptability rates for Image Quality, Operational Experience, and Stability were 99.3% (95% Bootstrap CI: 98.5% - 100.0%), 99.3% (95% Bootstrap CI: 98.3% - 100.0%), and 98.7% (95% Bootstrap CI: 97.0% - 100.0%), respectively, all exceeding the pre-defined performance goal of 84.4%. Secondary analysis revealed that the median scores for most items fell within the "High/Excellent" range (e.g., Resolution: 3 [IQR: 2-3]), indicating a generally positive perception. Detailed ordinal distributions highlighted that for image quality, focus and noise received the highest proportion of "Excellent" (Accurate/Low) ratings (89.5% and 81.6% respectively); for operational experience, button comfort rated best (65.8% Excellent); for stability, changes in device or probe performance rated best (85.5% Stable).

Conclusion

The Color Doppler Ultrasound System demonstrates robust safety and high performance in real-world clinical settings, meeting the pre-specified performance criteria. Secondary analyses provided granular insights into specific strengths and minor areas for future optimization.

Zitieren

APA

Wang, Y., Wang, F., Wu, D., Ma, J., Zhang, B., & Xie, W. (2026). REAL-WORLD SAFETY AND PERFORMANCE OF THE COLOR DOPPLER ULTRASOUND SYSTEM: A MULTI-CENTER POST-MARKET CLINICAL FOLLOW-UP SURVEY. Journal of Pharmaceutical and Medical Research. https://doi.org/10.61784/jpmr3093

Vancouver

Wang Y, Wang F, Wu D, Ma J, Zhang B, Xie W. REAL-WORLD SAFETY AND PERFORMANCE OF THE COLOR DOPPLER ULTRASOUND SYSTEM: A MULTI-CENTER POST-MARKET CLINICAL FOLLOW-UP SURVEY. Journal of Pharmaceutical and Medical Research. 2026. doi:10.61784/jpmr3093

BibTeX
@article{Wang2026,
  author = {Wang, YuanBo and Wang, FuCheng and Wu, Dan and Ma, JieBing and Zhang, BaoLi and Xie, Wen},
  title = {REAL-WORLD SAFETY AND PERFORMANCE OF THE COLOR DOPPLER ULTRASOUND SYSTEM: A MULTI-CENTER POST-MARKET CLINICAL FOLLOW-UP SURVEY},
  journal = {Journal of Pharmaceutical and Medical Research},
  year = {2026},
  doi = {10.61784/jpmr3093}
}
Kostenlos starten →