Abstract (Originalsprache)

Purpose

To compare the mean keratometry (K) readings and corneal astigmatism obtained from gold standard Manual keratometer with Automated keratometer, ORBSCAN II and IOLMaster 500 in patients undergoing Femto-second Laser Assisted Cataract Surgery (FLACS).

Methods

This prospective, non-randomized, comparative clinical study was done in 70 patients. A total of 120 eyes were included, each of which underwent keratometry with four different types of keratometers. Mean K values and corneal astigmatism were obtained from each machine and compared. BlandAltman plots were used to determine the 95% limits of agreement (LoA) between different devices.

Results

The Average K value from manual keratometer was 43.695±1.884(SD). Mean K from Auto-K was 43.686±1.917, ORBSCAN II was 43.748±1.865 and IOLMaster 500 was 43.685±1.875. Thus, keratometric values from all three newer devices correlated significantly (p ≤ 0.01) with the gold standard manual keratometer. In addition, the magnitude of corneal astigmatism measured by all three devices correlated significantly (p ≤ 0.01) with the gold standard manual keratometer.

Conclusion

The newer keratometers including Automated Keratometer, ORBSCAN II and IOLMaster 500 were found to be comparable to the gold standard Manual Keratometer. The keratometric readings and corneal astigmatism from each of the newer keratometers showed high degree of correlation with that of manual keratometer.

Zitieren

APA

Kumar, K. C., Tyagi, S., & Aggarwal, S. (2026). Inter-Device Agreement of Keratometric and Astigmatic Measurements Between Manual and Modern Keratometers pre-FLACS. Journal of Optometry and Ophthalmology Research, 1-6. https://doi.org/10.61440/joor.2026.v2.09

Vancouver

Kumar KC, Tyagi S, Aggarwal S. Inter-Device Agreement of Keratometric and Astigmatic Measurements Between Manual and Modern Keratometers pre-FLACS. Journal of Optometry and Ophthalmology Research. 2026:1-6. doi:10.61440/joor.2026.v2.09

BibTeX
@article{Kumar2026,
  author = {Kumar, Kritika Chopra and Tyagi, Swati and Aggarwal, Surbhi},
  title = {Inter-Device Agreement of Keratometric and Astigmatic Measurements Between Manual and Modern Keratometers pre-FLACS},
  journal = {Journal of Optometry and Ophthalmology Research},
  year = {2026},
  pages = {1--6},
  doi = {10.61440/joor.2026.v2.09}
}
Kostenlos starten →