Design and verification of the testing device for thoracic aortic stent grafts

Yu Zhou, Shiju Yan, Ailing Zhang

Progress in Medical Devices · 2025

DOI: 10.61189/063815kuibzu ↗

Kurz gefasst

Fachartikel in der Zeitschrift Progress in Medical Devices, 2025. Verfasst von Yu Zhou und 2 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To design a testing device for measuring the radial support force and bending spring back force of stent grafts and evaluate its effectiveness.

Methods

A radial force and spring-back force testing device was designed to integrate with a tensile testing machine. The radial compression and bending characteristics of stent grafts for thoracic aorta applications were analyzed, and the corresponding conversion formulas were derived. A custom stent ring fixture was fabricated, and a five-wave gradient stent was sewn. Both physical experiments and finite element simulations were conducted. Radial support forces were measured by gripping 20%, 40%, and 60% of the stent’s diameter, and bending tests were performed at angles of 60°, 90°, and 180°. The stability of the testing device was analyzed through comparative tests across different compression diameters and bending angles.

Results

The device demonstrated high detection precision, stability, and accuracy, with minimal deviation across multiple measurements. The mechanical behavior of the stent observed in both finite element simulations and physical experiments showed consistent results.

Conclusions

The testing device developed in this study effectively measures the mechanical changes in large-diameter stent grafts, providing a new reference for testing large-diameter stents.

Zitieren

APA

Zhou, Y., Yan, S., & Zhang, A. (2025). Design and verification of the testing device for thoracic aortic stent grafts. Progress in Medical Devices. https://doi.org/10.61189/063815kuibzu

Vancouver

Zhou Y, Yan S, Zhang A. Design and verification of the testing device for thoracic aortic stent grafts. Progress in Medical Devices. 2025. doi:10.61189/063815kuibzu

BibTeX
@article{Zhou2025,
  author = {Zhou, Yu and Yan, Shiju and Zhang, Ailing},
  title = {Design and verification of the testing device for thoracic aortic stent grafts},
  journal = {Progress in Medical Devices},
  year = {2025},
  doi = {10.61189/063815kuibzu}
}
Kostenlos starten →