Physical Properties of Small Focal Spot Imaging with Deep Learning Reconstruction in Chest-abdominal Plain CT
Japanese Journal of Radiological Technology · 2023
DOI: 10.6009/jjrt.2023-1400 ↗Kurz gefasst
Fachartikel in der Zeitschrift Japanese Journal of Radiological Technology, 2023. Verfasst von Makoto Fujiwara und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, JA Oita Koseiren Tsurumi Hospital. Hauptthemen laut MEDLINE: Deep Learning. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 35 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: The aim of this study was to compare the physical properties of small focal spot imaging with deep learning reconstruction (DLR) and small or large focal spot imaging with hybrid iterative reconstruction (IR) in chest-abdominal plain computed tomography. METHOD: In small focal spot imaging using DLR and hybrid IR, tube currents were set at 350 mA. For the large focal spot imaging using hybrid IR, the tube current was set at 360, 400, 450, and 500 mA. The spatial frequencies with 50% task transfer function (TTF) for delrin and acrylic were calculated to compare spatial resolution properties for lung and soft tissue in the chest. Additionally, the low-contrast object-specific contrast-to-noise ratio (CNRLO) was measured as noise property was measured for a 7-mm module with a CT value contrast of 10 HU in the abdomen. RESULT: Spatial frequencies with 50% TTF for delrin and acrylic were found to be greater in small focal spot imaging using DLR compared to those in small and large focal spot imaging using hybrid IR. Moreover, the CNRLO obtained from small focal spot imaging with DLR was also nearly equivalent to that of large focal spot imaging with hybrid IR at tube currents of 450 and 500 mA. CONCLUSION: In chest-abdominal plain computed tomography, small focal spot imaging with DLR has been demonstrated to exhibit greater spatial resolution properties compared to small and large focal spot imaging with hybrid IR, with equivalent or better noise performance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, JA Oita Koseiren Tsurumi Hospital.
- Medical Technology Department, Oita University Hospital.
- Canon Medical Systems Corporation.
Zitieren
APA
Fujiwara, M., Shiotsuki, K., Kawano, M., Noto, D., Maruyama, K., & Miyazaki, M. (2023). Physical Properties of Small Focal Spot Imaging with Deep Learning Reconstruction in Chest-abdominal Plain CT. Japanese Journal of Radiological Technology. https://doi.org/10.6009/jjrt.2023-1400
Vancouver
Fujiwara M, Shiotsuki K, Kawano M, Noto D, Maruyama K, Miyazaki M. Physical Properties of Small Focal Spot Imaging with Deep Learning Reconstruction in Chest-abdominal Plain CT. Japanese Journal of Radiological Technology. 2023. doi:10.6009/jjrt.2023-1400
BibTeX
@article{Fujiwara2023,
author = {Fujiwara, Makoto and Shiotsuki, Kenshi and Kawano, Mizuki and Noto, Daichi and Maruyama, Kenta and Miyazaki, Misaki},
title = {Physical Properties of Small Focal Spot Imaging with Deep Learning Reconstruction in Chest-abdominal Plain CT},
journal = {Japanese Journal of Radiological Technology},
year = {2023},
doi = {10.6009/jjrt.2023-1400}
}