EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

A Novel Electrochemical Sensor Based on MnO2/Sepiolite Nanocomposite for the Detection of Hydrogen Peroxide in Human Serum Samples

AbduRahman Hosseinifar, Masoud Ghanei-Motlagh, Maryam Fayazi

Annals of Military and Health Sciences Research · 23.11.2021 · 19 (3)

DOI: 10.5812/amh.119100 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Annals of Military and Health Sciences Research, online erschienen am 23.11.2021. Verfasst von AbduRahman Hosseinifar und 2 weiteren Autor:innen. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The reliable and easy-to-operate detection of hydrogen peroxide (H2O2) has attracted extensive attention in the fields of biomedicine, food security, and environmental analysis.

Objectives

In this work, a novel electrochemical method was proposed for H2O2 monitoring using a carbon paste electrode (CPE) modified with MnO2/sepiolite nanocomposite.

Methods

MnO2/sepiolite material was characterized by transmission electron microscopy (TEM), energy-dispersive X-ray spectroscopy (EDS), and X-ray diffraction (XRD) technique. The modified CPE was employed for the amperometric monitoring of H2O2 in human serum samples.

Results

Electrochemical data showed that the MnO2/sepiolite-CPE displays a high peak current towards H2O2 oxidation. A linear range from 5 to 700 μM and a low detection limit of 0.8 μM for H2O2 were obtained with the proposed sensor. Besides, the electrode depicted excellent reproducibility and anti-interferant ability, promising the applicability of this electrochemical method in practical analyses.

Conclusions

This work introduced a new and effective enzyme-less H2O2 sensor based on the MnO2/sepiolite nanocomposite modified CPE. The suggested sensor showed good sensitivity for the rapid detection of H2O2 in a wide linear range with a low detection limit and satisfactory reproducibility, which made it practical for the analysis of hydrogen H2O2 in real samples.

Zitieren

APA

Hosseinifar, A., Ghanei-Motlagh, M., & Fayazi, M. (2021). A Novel Electrochemical Sensor Based on MnO2/Sepiolite Nanocomposite for the Detection of Hydrogen Peroxide in Human Serum Samples. Annals of Military and Health Sciences Research, 19(3). https://doi.org/10.5812/amh.119100

Vancouver

Hosseinifar A, Ghanei-Motlagh M, Fayazi M. A Novel Electrochemical Sensor Based on MnO2/Sepiolite Nanocomposite for the Detection of Hydrogen Peroxide in Human Serum Samples. Annals of Military and Health Sciences Research. 2021;19(3). doi:10.5812/amh.119100

BibTeX
@article{Hosseinifar2021,
  author = {Hosseinifar, AbduRahman and Ghanei-Motlagh, Masoud and Fayazi, Maryam},
  title = {A Novel Electrochemical Sensor Based on MnO2/Sepiolite Nanocomposite for the Detection of Hydrogen Peroxide in Human Serum Samples},
  journal = {Annals of Military and Health Sciences Research},
  year = {2021},
  volume = {19},
  number = {3},
  doi = {10.5812/amh.119100}
}
Kostenlos starten →