EUVIMEDEUVIMED entdecken
FallberichtOpen Access

Post-Detachment Displacement of a Woven EndoBridge (WEB) Device Managed via Microcatheter Manipulation: A Case Report

Kyoichi Seo, Yoji Kuramoto, Soichiro Abe, Shinichi Yoshimura

Journal of Neuroendovascular Therapy · 2026 · 20 (1) · n/a

DOI: 10.5797/jnet.cr.2026-0107 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Journal of Neuroendovascular Therapy, 2026. Verfasst von Kyoichi Seo und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurosurgery, Hyogo Medical University, Nishinomiya, Hyogo, Japan. Die Arbeit nennt 13 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To report a case of post-detachment Woven EndoBridge (WEB) displacement successfully managed through microcatheter manipulation alone, avoiding unplanned stenting, which necessitates long-term dual antiplatelet therapy. Case Presentation: A 42-year-old male presented with an unruptured middle cerebral artery bifurcation aneurysm. Device sizing was planned using high-resolution 3D DSA and volumetric analysis on a Siemens ARTIS icono biplane (ARTIS icono VE20 software; Siemens Healthineers, Erlangen, Germany). A WEB SL 5 × 3 mm device (Terumo Neuro, Tustin, CA, USA) was deployed and detached. However, the subsequent withdrawal of an adjunctive microcatheter from a branch vessel caused vascular recoil, leading the detached WEB to rotate and protrude toward the branch orifice, compromising distal flow. Instead of rescue stenting, we performed a "pushing technique" using a Leonis Mova microcatheter (SB-Kawasumi Laboratories, Kanagawa, Japan). The device was successfully reseated within the aneurysm sac, albeit with a 90° rotation. Final angiography confirmed stable seating and excellent branch patency. The patient was discharged without neurological deficits on single antiplatelet therapy.

Conclusion

Post-detachment WEB displacement can occur due to dynamic vascular shifts even with precise volumetric sizing. Microcatheter manipulation is a feasible alternative bailout option that preserves the low-profile advantages of the WEB device and avoids the risks associated with rescue stenting.

Schlagwörter

Stichwörter der Autor:innen

Woven EndoBridge (WEB)bailout strategydevice displacementendovascular treatmentintracranial aneurysm

Autor:innen und Institutionen

  • Department of Neurosurgery, Hyogo Medical University, Nishinomiya, Hyogo, Japan.

Zitieren

APA

Seo, K., Kuramoto, Y., Abe, S., & Yoshimura, S. (2026). Post-Detachment Displacement of a Woven EndoBridge (WEB) Device Managed via Microcatheter Manipulation: A Case Report. Journal of Neuroendovascular Therapy, 20(1), n/a. https://doi.org/10.5797/jnet.cr.2026-0107

Vancouver

Seo K, Kuramoto Y, Abe S, Yoshimura S. Post-Detachment Displacement of a Woven EndoBridge (WEB) Device Managed via Microcatheter Manipulation: A Case Report. Journal of Neuroendovascular Therapy. 2026;20(1):n/a. doi:10.5797/jnet.cr.2026-0107

BibTeX
@article{Seo2026,
  author = {Seo, Kyoichi and Kuramoto, Yoji and Abe, Soichiro and Yoshimura, Shinichi},
  title = {Post-Detachment Displacement of a Woven EndoBridge (WEB) Device Managed via Microcatheter Manipulation: A Case Report},
  journal = {Journal of Neuroendovascular Therapy},
  year = {2026},
  volume = {20},
  number = {1},
  pages = {n/a},
  doi = {10.5797/jnet.cr.2026-0107}
}