EUVIMEDEUVIMED entdecken
Open Access

Use of intracardiac echocardiography and three-dimensional mapping during catheter ablation for atrial fibrillation is associated with reduced complications: a retrospective analysis of United States Medicare Fee-For-Service Database

Yiran Rong iD, Tom Wei, Kevin Corriveau, Rahul Khanna iD

Journal of Comparative Effectiveness Research · 2026

DOI: 10.57264/cer-2026-0037 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Comparative Effectiveness Research, 2026. Verfasst von Yiran Rong und 3 weiteren Autor:innen. Beteiligte Einrichtung: MedTech Epidemiology & Real-World Data Sciences, Johnson & Johnson, New Brunswick, NJ 08933, USA. Hauptthemen laut MEDLINE: Atrial Fibrillation, Catheter Ablation, Postoperative Complications, Echocardiography. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aim

This retrospective cohort study assessed differences in procedure-related complications and death among old patients with atrial fibrillation (AF) undergoing catheter ablation (CA) procedure with versus without use of intracardiac echocardiography (ICE) and advanced three-dimensional (3D) electroanatomical mapping. Materials & methods: Using the Medicare Database, a retrospective cohort of AF patients ≥65 years of age who underwent CA procedure were identified, and categorized into two cohorts: CA without ICE/3D mapping versus CA with ICE/3D mapping. Outcomes including a composite of complications (any), cardiac perforation, esophageal fistula and death, respectively, in the 30-day period post-CA were assessed. Inverse probability of treatment weighting method was used for covariate balancing. Cox regression models were performed to evaluate outcomes.

Results

There were 35,490 patients in CA without ICE/3D mapping cohort and 170,762 in the CA with ICE/3D mapping cohort. Cox regression model revealed that patients who underwent CA without ICE/3D mapping use were ∼3.5-times more likely to have 30-day any complication versus those who had CA with ICE/3D mapping (7.3% vs 2.2%, log-rank test p < 0.001; hazard ratio: 3.507; 95% CI: 3.307–3.720, p < 0.001). The incidence of cardiac perforation (2.5% vs 0.1%, log-rank test p < 0.001), atrio-esophageal fistula (0.03% vs 0.01%, log-rank test p = 0.03) and death (1.6% vs 0.2%, log-rank test p < 0.001) were significantly higher among patients without ICE/3D mapping versus those who had CA with ICE/3D mapping.

Conclusion

The absence of ICE and 3D mapping during CA procedures was associated with significantly increased risk of complications and death among Medicare eligible patients with AF.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Atrial FibrillationCatheter AblationPostoperative ComplicationsEchocardiographyHumansUnited StatesRetrospective StudiesFemaleMaleMedicareAgedAged, 80 and overDatabases, Factual

Stichwörter der Autor:innen

3D electroanatomical mappingatrial fibrillationcatheter ablationintracardiac echocardiography

Autor:innen und Institutionen

  • MedTech Epidemiology & Real-World Data Sciences, Johnson & Johnson, New Brunswick, NJ 08933, USA.
  • Health Economics & Market Access, Johnson & Johnson MedTech Electrophysiology Irvine, CA 92618, USA.

ORCID-Profile: Yiran Rong, Rahul Khanna

Zitieren

APA

Rong, Y., Wei, T., Corriveau, K., & Khanna, R. (2026). Use of intracardiac echocardiography and three-dimensional mapping during catheter ablation for atrial fibrillation is associated with reduced complications: a retrospective analysis of United States Medicare Fee-For-Service Database. Journal of Comparative Effectiveness Research. https://doi.org/10.57264/cer-2026-0037

Vancouver

Rong Y, Wei T, Corriveau K, Khanna R. Use of intracardiac echocardiography and three-dimensional mapping during catheter ablation for atrial fibrillation is associated with reduced complications: a retrospective analysis of United States Medicare Fee-For-Service Database. Journal of Comparative Effectiveness Research. 2026. doi:10.57264/cer-2026-0037

BibTeX
@article{Rong2026,
  author = {Rong, Yiran and Wei, Tom and Corriveau, Kevin and Khanna, Rahul},
  title = {Use of intracardiac echocardiography and three-dimensional mapping during catheter ablation for atrial fibrillation is associated with reduced complications: a retrospective analysis of United States Medicare Fee-For-Service Database},
  journal = {Journal of Comparative Effectiveness Research},
  year = {2026},
  doi = {10.57264/cer-2026-0037}
}
Kostenlos starten →