Interaction of positive airway pressure mask magnets with cardiac implantable electronic devices
Journal of Clinical Sleep Medicine · 2023-05 · 19 (5) · 941-946
DOI: 10.5664/jcsm.10478 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Sleep Medicine, 2023-05. Verfasst von Chad M Ruoff und 11 weiteren Autor:innen. Beteiligte Einrichtung: Division of Pulmonary Medicine, Mayo Clinic Hospital, Phoenix, Arizona. Hauptthemen laut MEDLINE: Magnets, Sleep Apnea Syndromes. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
STUDY OBJECTIVES: To evaluate for potential interactions between magnetic positive airway pressure (mPAP) masks and cardiac implantable electronic devices (CIEDs) for patients with sleep apnea. METHODS: Adult patients with a CIED who used an mPAP mask were recruited from our sleep clinic to undergo a safety visit at our pacemaker clinic. We tested whether the mPAP interacted with the implanted device at home during normal use and in the clinic during simulated normal use and with direct contact. The magnetic field strength of 6 mPAP masks was tested with a gaussmeter. RESULTS: Of 13 patients tested, 1 (8%), wearing a full face mask (ResMed AirFit F30 [ResMed, San Diego, California]), had a magnet response event (interaction) with direct contact, but no interactions were identified during normal or simulated normal use in any patient. The magnetic field strength of the mPAP masks increased the closer the mask got to the CIED, from 0.4 mT (4 G) at the mask manufacturer's recommended 5.1-cm (2-inch) distance from an implanted medical device up to 291 mT (2,910 G) at 0 cm (0 inches; direct contact). CONCLUSIONS: An mPAP mask may interact with a CIED if placed directly on the skin overlying the CIED. The use of Philips Respironics (Philips, Cambridge, Massachusetts) mPAP masks is now contraindicated in patients with a CIED. Until additional studies are conducted to better document the risks and benefits of mPAP masks, we recommend discouraging patients with CIEDs from using any mPAP mask. CITATION: Ruoff CM, Tashman YS, Cheema KPK, et al. Interaction of positive airway pressure mask magnets with cardiac implantable electronic devices. J Clin Sleep Med. 2023;19(5):941-946.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Pulmonary Medicine, Mayo Clinic Hospital, Phoenix, Arizona.
- Division of Heart Rhythm Services, Mayo Clinic School of Graduate Medical Education, Mayo Clinic College of Medicine and Science, Phoenix, Arizona.
- Division of Pulmonary Medicine, Mayo Clinic, Scottsdale, Arizona.
- Department of Cardiovascular Diseases, Mayo Clinic Hospital, Phoenix, Arizona.
Zitieren
APA
Ruoff, C. M., Tashman, Y. S., Cheema, K. P. K., Miller, B. W., Houser, R. F., Lizak, M. J., Haley, C. C., Petersen, A. M., Goswami, U., Lyng, P. J., Krahn, L. E., & Srivathsan, K. (2023). Interaction of positive airway pressure mask magnets with cardiac implantable electronic devices. Journal of Clinical Sleep Medicine, 19(5), 941-946. https://doi.org/10.5664/jcsm.10478
Vancouver
Ruoff CM, Tashman YS, Cheema KPK, Miller BW, Houser RF, Lizak MJ, et al. Interaction of positive airway pressure mask magnets with cardiac implantable electronic devices. Journal of Clinical Sleep Medicine. 2023;19(5):941-946. doi:10.5664/jcsm.10478
BibTeX
@article{Ruoff2023,
author = {Ruoff, Chad M and Tashman, Yasemin S and Cheema, Kamal Preet K and Miller, Bernie W and Houser, Ryan F and Lizak, Matthew J and Haley, Caitlin C and Petersen, Ann M and Goswami, Umesh and Lyng, Philip J and Krahn, Lois E and Srivathsan, Komandoor},
title = {Interaction of positive airway pressure mask magnets with cardiac implantable electronic devices},
journal = {Journal of Clinical Sleep Medicine},
year = {2023},
volume = {19},
number = {5},
pages = {941--946},
doi = {10.5664/jcsm.10478}
}
EUVIMED