Feasibility of watch-based blood pressure monitoring device in daily blood pressure monitoring
Clinical Hypertension · 2025 · 31 (1)
DOI: 10.5646/ch.2025.31.e21 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Hypertension, 2025. Verfasst von Youngro Lee und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical and Computer Engineering, Seoul National University College of Engineering, Seoul, Republic of Korea. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: National Research Foundation of Korea.
Abstract (Originalsprache)
Background
Cuffless blood pressure (BP) measurement devices integrated into smartwatches have gained prominence, yet limited studies provide the feasibility and preciseness of daily BP monitoring. Here, we evaluated the trackability of daily BP variance and the precision of the calibration process.
Methods
We collected the data from 896 participants, reporting 35,592 BP values, and body composition analysis data measured by the Samsung Galaxy Watch 6 device. Participants were instructed to measure BP daily, in the morning (5 AM-9 AM) and evening (6 PM-10 PM) for 2 weeks, with initial calibration and re-calibration after the first week. Body composition data, obtained using the Galaxy Watch's bioelectrical impedance analysis sensor, was measured voluntarily during the campaign without specific time constraints.
Results
With BP readings collected using smartwatches, morning and evening BP values showed a significant difference, higher in the evening by 1.42 ± 5.25 mmHg (P < 0.05). Basal metabolic rate, skeletal muscle mass, total body water, morning systolic BP, morning pulse pressure, and morning heart rate were significantly associated with higher difference in morning-evening BP. The calibration stability was assessed by the difference in average BP before and after calibration, showing a substantial pre-post calibration BP difference by 4.64 ± 4.73 mmHg of systolic BP and 3.66 ± 3.62 mmHg of diastolic BP.
Conclusions
In conclusion, watch-based devices may not detect clinical-level BP variability, and substantial extent of pre-post calibration error has to be solved for their utility in regular real-life BP monitoring.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Electrical and Computer Engineering, Seoul National University College of Engineering, Seoul, Republic of Korea.
- Institute of Engineering Research at Seoul National University, Seoul, Republic of Korea.
- Inter-university Semiconductor Research Center (ISRC), Seoul National University, Seoul, Republic of Korea.
- Department of Internal Medicine, Seoul National University College of Medicine, Seoul, Republic of Korea.
- Department of Internal Medicine, Seoul National University Hospital, Seoul, Republic of Korea.
- Biomedical Research Institute, Seoul National University Hospital, Seoul, Republic of Korea.
ORCID-Profile: Youngro Lee, Sungjoon Park, Jongae Park, Jongmo Seo, Hae-Young Lee
Förderung
- National Research Foundation of Korea
Zitieren
APA
Lee, Y., Park, S., Park, J., Seo, J., & Lee, H. Y. (2025). Feasibility of watch-based blood pressure monitoring device in daily blood pressure monitoring. Clinical Hypertension, 31(1). https://doi.org/10.5646/ch.2025.31.e21
Vancouver
Lee Y, Park S, Park J, Seo J, Lee HY. Feasibility of watch-based blood pressure monitoring device in daily blood pressure monitoring. Clinical Hypertension. 2025;31(1). doi:10.5646/ch.2025.31.e21
BibTeX
@article{Lee2025,
author = {Lee, Youngro and Park, Sungjoon and Park, Jongae and Seo, Jongmo and Lee, Hae-Young},
title = {Feasibility of watch-based blood pressure monitoring device in daily blood pressure monitoring},
journal = {Clinical Hypertension},
year = {2025},
volume = {31},
number = {1},
doi = {10.5646/ch.2025.31.e21}
}
EUVIMED