Differences in Cholesterol Test Results Between POCT (Point of Care Testing) and Photometer Methods
INDOGENIUS · 01.02.2026 · 5 (1) · 31-36
DOI: 10.56359/igj.v5i1.671 ↗Kurz gefasst
Fachartikel in der Zeitschrift INDOGENIUS, online erschienen am 01.02.2026. Verfasst von Suparyati und 1 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background & Objective: Cholesterol is one of the components of fat or lipids that is essential for the body, in addition to carbohydrates, proteins, vitamins, and minerals. Cholesterol testing in the laboratory can be performed using two types of methods, namely automatic technique testing (POCT) and simple technique testing (photometer). This study aims to determine the difference in cholesterol test results using the POCT method and the photometer method.
Method
This study is analytical in nature, using 30 venous blood samples from outpatients at Kraton Pekalongan Regional General Hospital. The examination was conducted using the POCT method with a photometer. The data were analyzed using normality tests and Wilcoxon tests using SPSS.
Result
The average cholesterol test result using the POCT method was 173.37 mg/dl, while the photometer method yielded a result of 127.77 mg/dl. The statistical test results showed (sig value = 0.000 < 0.05), indicating a significant difference between the two methods.
Conclusion
There is a significant difference between the cholesterol results obtained by the POCT method and the photometer.
Zitieren
APA
Suparyati, & Anggraini, A. C. (2026). Differences in Cholesterol Test Results Between POCT (Point of Care Testing) and Photometer Methods. INDOGENIUS, 5(1), 31-36. https://doi.org/10.56359/igj.v5i1.671
Vancouver
Suparyati, Anggraini AC. Differences in Cholesterol Test Results Between POCT (Point of Care Testing) and Photometer Methods. INDOGENIUS. 2026;5(1):31-36. doi:10.56359/igj.v5i1.671
BibTeX
@article{Suparyati2026,
author = {Suparyati and Anggraini, Adelia Citra},
title = {Differences in Cholesterol Test Results Between POCT (Point of Care Testing) and Photometer Methods},
journal = {INDOGENIUS},
year = {2026},
volume = {5},
number = {1},
pages = {31--36},
doi = {10.56359/igj.v5i1.671}
}