The influence of the negative pressure value on the efficiency of drainage of exudates of various densities in negative pressure therapy – an experimental model
Polish Journal of Surgery · 14.04.2025 · 97 (3) · 63-70
DOI: 10.5604/01.3001.0055.0863 ↗Kurz gefasst
Fachartikel in der Zeitschrift Polish Journal of Surgery, erschienen am 14.04.2025. Verfasst von Maciej Okła und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department and Clinic of Maxillofacial Surgery, K. Marcinkowski Medical University in Poznan, Poland. Hauptthemen laut MEDLINE: Negative-Pressure Wound Therapy, Exudates and Transudates, Drainage, Wound Healing. Die Arbeit nennt 24 Literaturangaben.
Abstract (Originalsprache)
Introduction
Negative pressure therapy is used to promote healing in many types of wounds at various stages of healing. Little is known about the effect of the value of the applied negative pressure on the efficiency of exudate drainage, particularly with regard to exudates of different densities.
Aim
The aim of this study was to evaluate the effectiveness of drainage of exudates of different densities using negative pressures of −80 mmHg and −120 mmHg, without or with the use of an intermediate layer.
Material and method: An experimental craniofacial model was developed and formed of tissue-like silicone. Three identical models were used simultaneously for study purposes. Wound imitations were made in the models in a manner facilitating the delivery of fluids of different densities to the wound bottom. After the vacuum dressing was applied, low-density fluid (water, A), medium-density fluid (filtered 30% cream, B) and high-density fluid (unfiltered 30% cream, C) used as a model of purulent secretions, were administered sequentially; in the cases of low-density and high-density fluids, the delivery was repeated with the use of an intermediate layer (D, E). Each time, the fluids were administered in the quantity of 100 mL, and the assessments were made at negative pressure values of −80 mmHg and −120 mmHg by measuring the time between the onset of delivery and complete discharge of secretions.
Results
For low- and medium-density fluids without an intermediate layer (A and B), no differences in its drainage time were observed for pressures of −80 mmHg and −120 mmHg. For the high-density fluid (C), drainage times were significantly longer than in the case of low- and medium-density fluids. At −120 mmHg, the drainage time for high-density fluid was statistically significantly shorter than at −80 mmHg. When an intermediate layer was used for low-density fluid (D) no differences were observed between both pressure values as well as in the comparison to the model without an intermediate layer. When an intermediate layer was used for high-density fluid (E), the drainage time was significantly longer than in the model without the intermediate layer (C). In the tests using dense fluid and an intermediate layer, no significant difference in drainage time was observed for both pressure values.
Conclusions
The type of secretion, and particularly its density, should be taken into account when selecting the value of vacuum applied in negative pressure therapy; higher pressures should be used for dense secretions. For dense secretions, drainage becomes difficult upon the presence of an intermediate layer, regardless of the pressure applied.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department and Clinic of Maxillofacial Surgery, K. Marcinkowski Medical University in Poznan, Poland.
- K. Marcinkowski Medical University in Poznan, Poland.
- Department of General and Oncological Surgery, St. Padre Pio Regional Hospital in Przemysl, Poland.
- Department of Oncological Surgery, Jędrzej Śniadecki Specialist Hospital in Nowy Sacz, Poland.
- Department of Conservative Dentistry and Endodontics, K. Marcinkowski Medical University of Poznan, Poland.
- Department of General, Endocrine Surgery and Gastroenterological Oncology, K. Marcinkowski Medical University of Poznan, Poland.
Zitieren
APA
Okła, M., Szczepaniak, M., Organ, A., Molin, M., Słowik, Ł., Nijakowski, K., & Banasiewicz, T. (2025). The influence of the negative pressure value on the efficiency of drainage of exudates of various densities in negative pressure therapy – an experimental model. Polish Journal of Surgery, 97(3), 63-70. https://doi.org/10.5604/01.3001.0055.0863
Vancouver
Okła M, Szczepaniak M, Organ A, Molin M, Słowik Ł, Nijakowski K, et al. The influence of the negative pressure value on the efficiency of drainage of exudates of various densities in negative pressure therapy – an experimental model. Polish Journal of Surgery. 2025;97(3):63-70. doi:10.5604/01.3001.0055.0863
BibTeX
@article{Oka2025,
author = {Okła, Maciej and Szczepaniak, Maria and Organ, Andrzej and Molin, Michał and Słowik, Łukasz and Nijakowski, Kacper and Banasiewicz, Tomasz},
title = {The influence of the negative pressure value on the efficiency of drainage of exudates of various densities in negative pressure therapy – an experimental model},
journal = {Polish Journal of Surgery},
year = {2025},
volume = {97},
number = {3},
pages = {63--70},
doi = {10.5604/01.3001.0055.0863}
}
EUVIMED