Evaluation of Transimpedance Matrix Measurement (TIM) heatmap for the assessment of cochlear implant electrode placement in patients with Inner Ear Malformations
Otolaryngologia Polska · 2025
DOI: 10.5604/01.3001.0055.0473 ↗Kurz gefasst
Fachartikel in der Zeitschrift Otolaryngologia Polska, 2025. Verfasst von Katarzyna Radomska und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otolaryngology, Pomeranian University of Medicine, Szczecin, Poland. Hauptthemen laut MEDLINE: Cochlear Implantation, Cochlear Implants, Ear, Inner, Acoustic Impedance Tests. Die Arbeit nennt 28 Literaturangaben.
Abstract (Originalsprache)
Introduction
It is estimated that 20% of pediatric patients eligible for implantation have various forms of Inner Ear Malformations (IEM). Transimpedance Matrix Measurement (TIM) is a relatively novel electrophysiological measurement protocol of the impedance patterns of electrode contacts within the cochlea.
Aim
Presentation of TIM measurements, in the form of a heatmap, in patients with IEM.
Materials and methods: Among 112 implantations performed between 2018 and 2024 in the pediatric group, 23 patients were diagnosed with IEM (20.5%). TIM and Stenver’s plain X-ray were performed for all patients.
Results
In 19 patients, standard approach with mastoidectomy and posterior tympanotomy was performed, in 2 cases retrofacial approach, in 2 petrosectomy with blind sac procedure, and in 2 “banana technique”. In most patients, electrode Contour Advance (CA) was used. One case of tip fold-over showed a characteristic cross pattern. For all other IEM there were normal diagonals but voltage distribution showed abnormal distribution within the cochlea.
Discussion
Most patients received a CA electrode because of its greater rigidity and the possibility of inhibiting CSF leakage by applying a piece of connective tissue for sealing to the electrode. Heatmaps for different IEM types differ from each other, but it is still possible to recognize TFO. The voltage spread in the cochlea in IEM shows different patterns for particular types of anomalies.
Conclusions
TIM measurement as a heatmap seems to be a quick and reliable method for detecting electrode array tip fold-over intraoperatively for patients with inner ear malformations. Our study complements the knowledge about the use of research based on TIM heatmap.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Otolaryngology, Pomeranian University of Medicine, Szczecin, Poland.
- Department of Otolaryngology, Pomeranian University of Medicine, Szczecin, Poland, Medicus Sp. z o.o., Wroclaw, Poland.
- Pediatric Cochlear Implant Program, Department of Otolaryngology, Audiology and Phoniatrics, Children's Hospital of Bydgoszcz, Bydgoszcz, Poland.
- Pediatric Cochlear Implant Program, Department of Otolaryngology, Audiology and Phoniatrics, Children's Hospital of Bydgoszcz, Bydgoszcz, Poland, Department of Developmental Age Diseases, Nicolaus Copernicus University,…
Zitieren
APA
Radomska, K., Talar, M., Tyra, J., & Mierzwiński, J. (2025). Evaluation of Transimpedance Matrix Measurement (TIM) heatmap for the assessment of cochlear implant electrode placement in patients with Inner Ear Malformations. Otolaryngologia Polska. https://doi.org/10.5604/01.3001.0055.0473
Vancouver
Radomska K, Talar M, Tyra J, Mierzwiński J. Evaluation of Transimpedance Matrix Measurement (TIM) heatmap for the assessment of cochlear implant electrode placement in patients with Inner Ear Malformations. Otolaryngologia Polska. 2025. doi:10.5604/01.3001.0055.0473
BibTeX
@article{Radomska2025,
author = {Radomska, Katarzyna and Talar, Marcin and Tyra, Justyna and Mierzwiński, Józef},
title = {Evaluation of Transimpedance Matrix Measurement (TIM) heatmap for the assessment of cochlear implant electrode placement in patients with Inner Ear Malformations},
journal = {Otolaryngologia Polska},
year = {2025},
doi = {10.5604/01.3001.0055.0473}
}
EUVIMED