Clinical usefulness of serum free light chains measurement in patients with multiple myeloma: comparative analysis of two different tests

Tadeusz Kubicki, Dominik Dytfeld, Aleksandra Baszczuk, Ewa Wysocka, Mieczysław Komarnicki, Krzysztof Lewandowski

Postępy Higieny i Medycyny Doświadczalnej · 2017

DOI: 10.5604/01.3001.0010.3788 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Postępy Higieny i Medycyny Doświadczalnej, 2017. Verfasst von Tadeusz Kubicki und 5 weiteren Autor:innen. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 18 Literaturangaben.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

There are two commercially available tests for measurement of serum free light chains (sFLC) in multiple myeloma (MM) patients – Freelite and N Latex FLC. The aim of this study was to perform an assessment and direct comparison of the usefulness of the methods in routine clinical practice.

Methods

40 refractory/relapsed MM patients underwent routine disease activity assessment studies, along with sFLC analysis using both assays. Correlation and concordance between the tests and sensitivity of studied methods of sFLC assessment were established. Special attention was focused on sFLC results in patients finally evaluated after completing the treatment.

Results

A weak correlation for the measurement of both κ [Passing–Bablok slope (PB) = 0.7681] and λ chains [(PB) = 1.542] was found. Using Bland–Altman plots, a bias of 0.0467 (κ) and -0.2133 (λ) between the measurements was documented. The concordance coefficient equaled 0.87 for κ, 0.62 for λ and 0.52 for κ/λ ratio. Ten patients had an abnormal Freelite assay κ/λ ratio and normal N Latex FLC κ/λ ratio. Three of these patients had negative serum protein electrophoresis results and fulfilled diagnostic criteria of stringent complete remission (sCR) according to N Latex FLC (but not according to Freelite). When the κ/λ ratio obtained by both methods was compared to patients’ serum/urine protein electrophoresis and immunofixation results, sensitivity of Freelite and N Latex FLC was established to be 62.5% and 41%, respectively.

Conclusions

There was no strong correlation or concordance between the two assays, and the sensitivity in terms of sFLC detection was different. This may cause problems when diagnosis of sCR is considered.

Zitieren

APA

Kubicki, T., Dytfeld, D., Baszczuk, A., Wysocka, E., Komarnicki, M., & Lewandowski, K. (2017). Clinical usefulness of serum free light chains measurement in patients with multiple myeloma: comparative analysis of two different tests. Postępy Higieny i Medycyny Doświadczalnej. https://doi.org/10.5604/01.3001.0010.3788

Vancouver

Kubicki T, Dytfeld D, Baszczuk A, Wysocka E, Komarnicki M, Lewandowski K. Clinical usefulness of serum free light chains measurement in patients with multiple myeloma: comparative analysis of two different tests. Postępy Higieny i Medycyny Doświadczalnej. 2017. doi:10.5604/01.3001.0010.3788

BibTeX
@article{Kubicki2017,
  author = {Kubicki, Tadeusz and Dytfeld, Dominik and Baszczuk, Aleksandra and Wysocka, Ewa and Komarnicki, Mieczysław and Lewandowski, Krzysztof},
  title = {Clinical usefulness of serum free light chains measurement in patients with multiple myeloma: comparative analysis of two different tests},
  journal = {Postępy Higieny i Medycyny Doświadczalnej},
  year = {2017},
  doi = {10.5604/01.3001.0010.3788}
}
Kostenlos starten →