Reliability of handheld autorefractometers in patients after cataract surgery
Turkish Journal of Medical Sciences · 14.10.2025 · 55 (5) · 1249-1255
DOI: 10.55730/1300-0144.6078 ↗Kurz gefasst
Fachartikel in der Zeitschrift Turkish Journal of Medical Sciences, online erschienen am 14.10.2025. Verfasst von Ayça Küpeli Çinar und 2 weiteren Autor:innen. Beteiligte Einrichtung: Division of Anterior Segment, Cornea and Refraction, Department of Ophthalmology, Faculty of Medicine, Trakya University, Edirne, Turkiye. Hauptthemen laut MEDLINE: Refraction, Ocular, Phacoemulsification, Cataract Extraction. Die Arbeit wurde bisher 1-mal zitiert (Crossref). Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background/aim: Accurate postoperative refraction assessment is essential for optimal visual rehabilitation after cataract surgery. Although handheld autorefractometers (HH-ARMs) are widely used in pediatric and uncooperative patients, their reliability in adults after cataract surgery remains uncertain. This study compared refractive outcomes obtained with a HH-ARM (Mediworks V100; Mediworks, Guangzhou, China) and a TB-ARM (Canon RK-F2; Canon Inc., Tokyo, Japan) to evaluate the clinical applicability of HH-ARM in this population. Materials and methods: This prospective study included 150 eyes of 150 patients (aged 55–85 years) who underwent uncomplicated phacoemulsification with intraocular lens (IOL) implantation. At 1 month postoperatively, spherical value (SV), cylindrical value (CV), cylindrical axis (CVA), and spherical equivalent (SE) were measured under cycloplegia with both devices. Agreement between devices was evaluated using nonparametric tests and Bland–Altman analysis.
Results
SV and CV differed significantly between devices (p = 0.04 and p < 0.001, respectively), while no significant differences were observed in CVA (p = 0.44) or SE (p = 0.09). Bland–Altman analysis confirmed close agreement for SE (mean difference –0.08 D; 95% limits of agreement [LoA] –1.57 D to 1.42 D). Regression analysis indicated that SE values were statistically interchangeable, whereas SV and CV were not.
Conclusion
HH-ARM demonstrated clinically acceptable agreement with TB-ARM for SE values, making SE the most reliable parameter for postoperative refraction in adults. These findings underscore the clinical relevance of HH-ARM as a practical alternative for patients with limited mobility and represent a novel contribution by validating its use in a population where evidence has been scarce. Further studies comparing HH-ARM with subjective refraction and across diverse surgical contexts are warranted.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Anterior Segment, Cornea and Refraction, Department of Ophthalmology, Faculty of Medicine, Trakya University, Edirne, Turkiye.
ORCID-Profile: Ayça Küpeli Çinar, Abdulkadir Can Çinar, Hande Güçlü
Zitieren
APA
Küpeli Çinar, A., Çinar, A. C., & Güçlü, H. (2025). Reliability of handheld autorefractometers in patients after cataract surgery . Turkish Journal of Medical Sciences, 55(5), 1249-1255. https://doi.org/10.55730/1300-0144.6078
Vancouver
Küpeli Çinar A, Çinar AC, Güçlü H. Reliability of handheld autorefractometers in patients after cataract surgery . Turkish Journal of Medical Sciences. 2025;55(5):1249-1255. doi:10.55730/1300-0144.6078
BibTeX
@article{Kpeliinar2025,
author = {Küpeli Çinar, Ayça and Çinar, Abdulkadir Can and Güçlü, Hande},
title = {Reliability of handheld autorefractometers in patients after cataract surgery },
journal = {Turkish Journal of Medical Sciences},
year = {2025},
volume = {55},
number = {5},
pages = {1249--1255},
doi = {10.55730/1300-0144.6078}
}
EUVIMED