10-Gauge versus 14-Gauge Fenestrated Needle/Catheter Units for Decompression of Tension Pneumothorax in Cadaveric Model
Journal of Special Operations Medicine · 2025
DOI: 10.55460/3luf-6pw2 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Special Operations Medicine, 2025. Verfasst von Adrianna N Long und 12 weiteren Autor:innen. Beteiligte Einrichtung: University of Texas Health Science Center San Antonio, TX. Hauptthemen laut MEDLINE: Pneumothorax, Needles, Decompression, Surgical, Catheters.
Abstract (Originalsprache)
BACKGROUND: Needle decompression (NDC) is the primary treatment of tension pneumothorax (tPTX) in prehospital settings. This study compared 10-gauge (10ga) and 14-gauge (14ga) fenestrated needle/catheter units for NDC. We hypoth-esized 10ga needle/catheter units would demonstrate higher tPTX decompression rates compared to 14ga needle/catheter units. METHODS: A non-randomized, non-blinded study was conducted using human cadavers with artificially induced tPTX (pleural pressure of 15mmHg). A 10ga or 14ga unit was in-serted into the 5th intercostal space, anterior axillary line, or the 2nd intercostal space, midclavicular line. Successful NDC was defined as a pressure decrease to less than 4mmHg. RESULTS: In 116 NDC attempts, there was no difference in the success rate of NDC between 10ga versus 14ga units (91.1% vs. 91.1%, P=1.0). The median time to decompression of tPTX was faster using 10ga at 22.0s (IQR 14.5-42.0) vs. 14ga at 39.8 seconds (IQR 30.3-57.6, P
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- University of Texas Health Science Center San Antonio, TX.
- Brooke Army Medical Center, San Antonio, TX.
- Center for Emergency Health Sciences, Spring Branch, TX.
- University of Texas Health Science Center San Antonio.
- ODP/Trauma Burn and Rehabilita-tive Medicine Team, U.S. Embassy, Abu Dhabi.
- 351st Special Warfare Training Squadron, Kirtland AFB, Albuquerque, NM.
Zitieren
APA
Long, A. N., Achay, J., Hudson, I. L., Lowe, J. B., Epley, E., Bolleter, S., DeSoucy, E. S., Hewitt, C. W., Rollman, J. E., Swenson, J. F., Wampler, D., Raetz, E., & Kruse, A. (2025). 10-Gauge versus 14-Gauge Fenestrated Needle/Catheter Units for Decompression of Tension Pneumothorax in Cadaveric Model. Journal of Special Operations Medicine. https://doi.org/10.55460/3luf-6pw2
Vancouver
Long AN, Achay J, Hudson IL, Lowe JB, Epley E, Bolleter S, et al. 10-Gauge versus 14-Gauge Fenestrated Needle/Catheter Units for Decompression of Tension Pneumothorax in Cadaveric Model. Journal of Special Operations Medicine. 2025. doi:10.55460/3luf-6pw2
BibTeX
@article{Long2025,
author = {Long, Adrianna N and Achay, Jennifer and Hudson, Ian L and Lowe, Joshua B and Epley, Emily and Bolleter, Scotty and DeSoucy, Erik Scott and Hewitt, Christopher W and Rollman, Jeffrey E and Swenson, Jeffrey F and Wampler, David and Raetz, Emily and Kruse, Adam},
title = {10-Gauge versus 14-Gauge Fenestrated Needle/Catheter Units for Decompression of Tension Pneumothorax in Cadaveric Model},
journal = {Journal of Special Operations Medicine},
year = {2025},
doi = {10.55460/3luf-6pw2}
}
EUVIMED