ANALISIS TINGKAT PAPARAN RADIASI DI AREA SETINGGI MATA DOKTER OPERATOR SELAMA PEMERIKSAAN DIAGNOSTIK CORONARY ANGIOGRAPHY DI RUANG CATHLAB
JRI (Jurnal Radiografer Indonesia) · 2018
DOI: 10.55451/jri.v1i1.6 ↗Kurz gefasst
Fachartikel in der Zeitschrift JRI (Jurnal Radiografer Indonesia), 2018. Verfasst von Devi Aprilia Primaningtyas und 2 weiteren Autor:innen.
Abstract (Originalsprache)
Background
The radiation exposure of interventional cardiology can provide biological effects on patients and workers that is deterministic and stochastic effects. One organ that accepts the possibility of deterministic effects are the eyes. To find out the possibility of cataracts is measuring the level of radiation exposure in the area of radiation workers at eye level during the diagnostic coronary angiography procedures. Method: the measurement of the level of radiation exposure in the area of physician eye-level using a survey meter directly at the physician position. Measurements were made on the variation of the projection during diagnostic coronary angiography examination including PA-Cranial, LAO-Cranial, LAO-Caudal, RAO-Cranial, RAO-Caudal, RAO and LAO.
Result
Measuring the level of radiation exposure in the area at physician eye-level during diagnostic coronary angiography obtained on sample A of 15.34 μSv for 69.17 seconds (1 minute 9.17 seconds and sample B of 25.11 μSv for 46 seconds. The total dose, DAP (dose Area Product) and fluoroscopy time of the operator A is higher than the operator B.
Conclusion
The level of radiation exposure in the area of physician eye-level during diagnostic coronary angiography examination obtained does not exceed the dose limit value (NBD) is 75 μSv /h.
Zitieren
APA
Primaningtyas, D. A., Praptono, S. I., & Muqmiroh, L. (2018). ANALISIS TINGKAT PAPARAN RADIASI DI AREA SETINGGI MATA DOKTER OPERATOR SELAMA PEMERIKSAAN DIAGNOSTIK CORONARY ANGIOGRAPHY DI RUANG CATHLAB. JRI (Jurnal Radiografer Indonesia). https://doi.org/10.55451/jri.v1i1.6
Vancouver
Primaningtyas DA, Praptono SI, Muqmiroh L. ANALISIS TINGKAT PAPARAN RADIASI DI AREA SETINGGI MATA DOKTER OPERATOR SELAMA PEMERIKSAAN DIAGNOSTIK CORONARY ANGIOGRAPHY DI RUANG CATHLAB. JRI (Jurnal Radiografer Indonesia). 2018. doi:10.55451/jri.v1i1.6
BibTeX
@article{Primaningtyas2018,
author = {Primaningtyas, Devi Aprilia and Praptono, Soegardo Indra and Muqmiroh, Lailatul},
title = {ANALISIS TINGKAT PAPARAN RADIASI DI AREA SETINGGI MATA DOKTER OPERATOR SELAMA PEMERIKSAAN DIAGNOSTIK CORONARY ANGIOGRAPHY DI RUANG CATHLAB},
journal = {JRI (Jurnal Radiografer Indonesia)},
year = {2018},
doi = {10.55451/jri.v1i1.6}
}