PENGARUH VARIASI ROTATION TIME TERHADAP NILAI DLP (DOSE LENGTH PRODUCT) DAN IMAGE NOISE PADA PESAWAT MSCT 128 SLICE (STUDI KASUS)

Asmaul Fauziyah, Amillia Kartika Sari, Soegardo Indra P, Anggraini Dwi Sensusiati, Pramono

JRI (Jurnal Radiografer Indonesia) · 2018

DOI: 10.55451/jri.v1i1.5 ↗

Kurz gefasst

Fachartikel in der Zeitschrift JRI (Jurnal Radiografer Indonesia), 2018. Verfasst von Asmaul Fauziyah und 4 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The patient's dose and image noise illustrate the quality of the CT scan. DLP (Dose Lenght Product) is a dose description on a CT-Scan tool by providing a measure of the total amount of radiation exposure for a series of scans. While image noise is the difference / variation in pixel values contained in an image matrix. The use of rotation time in CT scan is the most likely to contribute to patient dose and image noise.

Method

Toshiba Aquilion 128 MSCT aircraft and Acrilyc phantom were used as material in this study. The parameters on the CT-Scan tool used are the suitability test parameters, namely at kV 120, 400mA, slice thickness 2 mm, FOV 240. Rotation time variations on CT scan imaging that can be selected are in the range of values ​​of 0.35s, 0.375s, 0.4s , 0.45s, 0.5s, 0.6s, 0.75s, 1.0s while the other parameters are the same. From each rotation time value, the estimated dose value is the DLP value on the CT Scan monitor screen (Summary). Furthermore, the scanning results measure the noise value by measuring the ROI on the middle side of the area around a diameter of 2-3 cm2. The measurement results are recorded and entered into tables, graphs and tests with statistics, and test the effect (regression) as hypothesis testing with a significant value (α)

Zitieren

APA

Fauziyah, A., Sari, A. K., P, S. I., Sensusiati, A. D., & Pramono (2018). PENGARUH VARIASI ROTATION TIME TERHADAP NILAI DLP (DOSE LENGTH PRODUCT) DAN IMAGE NOISE PADA PESAWAT MSCT 128 SLICE (STUDI KASUS). JRI (Jurnal Radiografer Indonesia). https://doi.org/10.55451/jri.v1i1.5

Vancouver

Fauziyah A, Sari AK, P SI, Sensusiati AD, Pramono. PENGARUH VARIASI ROTATION TIME TERHADAP NILAI DLP (DOSE LENGTH PRODUCT) DAN IMAGE NOISE PADA PESAWAT MSCT 128 SLICE (STUDI KASUS). JRI (Jurnal Radiografer Indonesia). 2018. doi:10.55451/jri.v1i1.5

BibTeX
@article{Fauziyah2018,
  author = {Fauziyah, Asmaul and Sari, Amillia Kartika and P, Soegardo Indra and Sensusiati, Anggraini Dwi and Pramono},
  title = {PENGARUH VARIASI ROTATION TIME TERHADAP NILAI DLP (DOSE LENGTH PRODUCT) DAN IMAGE NOISE PADA PESAWAT MSCT 128 SLICE (STUDI KASUS)},
  journal = {JRI (Jurnal Radiografer Indonesia)},
  year = {2018},
  doi = {10.55451/jri.v1i1.5}
}
Kostenlos starten →