CONFORMITY TEST OF THE COLLIMATOR BEAM TO X-RAY ON DIAGNOSTIC X-RAY MACHINE AT THE RADIOLOGY INSTALLATION OF ARIFIN ACHMAD HOSPITAL, RIAU

Yoga Saputra, Marido Bisra

Medical Imaging and Radiation Protection Research (MIROR) Journal · 2021

DOI: 10.54973/miror.v1i1.59 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Imaging and Radiation Protection Research (MIROR) Journal, 2021. Verfasst von Yoga Saputra und 1 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Conformity test needs to be done in order to avoid deviation of X-rays photo. One of the test is conformity test collimator beam to X-ray. It aims to make sure angle and distance shift does not occur. This study was done at Radiology Installation of Arifin Achmad Hospital, Riau. This study is quantitative with experimental design by using collimator test tool. Tests were be done three times to each big and small focal spot. Based on Ministry of Health No. 1250 on 2009, deviation tolerance limit value is ≤ 2% FFD for conformity test of collimator beam and being ≤ 3° for the center point conformity test. The results of calculations on the small focal spot for X1 + X2 are respectively 2,25 cm, 2,25 cm, and 2,25 cm with an average value of 2,25 cm. In the calculation results of Y1 + Y2 for small focal spots, the results are respectively 0,45 cm, 0,50 cm and 0,50 cm with an average value of 0,48 cm. Then the calculation results on the large focal spot for X1 + X2 obtained results respectively 2,25 cm, 2,20 cm and 2,15 cm with an average value of 2,20 cm. the results of the calculation of Y1 + Y2 for large focal spots, respectively are 0,45 cm, 0,45 cm, and 0,45 cm with an average value of 0,45 cm. This conformity test got the deviation value exceed the tolerance limit on X axis namely of 2.25 cm in average on small focal spots and 2.20 cm on big focal spots. Conformity test results of the beam alignment entirely showed that center point was in a small circle with < 1.5° value. This means there is no deviation of center point which exceed tolerance limit.

Zitieren

APA

Saputra, Y., & Bisra, M. (2021). CONFORMITY TEST OF THE COLLIMATOR BEAM TO X-RAY ON DIAGNOSTIC X-RAY MACHINE AT THE RADIOLOGY INSTALLATION OF ARIFIN ACHMAD HOSPITAL, RIAU. Medical Imaging and Radiation Protection Research (MIROR) Journal. https://doi.org/10.54973/miror.v1i1.59

Vancouver

Saputra Y, Bisra M. CONFORMITY TEST OF THE COLLIMATOR BEAM TO X-RAY ON DIAGNOSTIC X-RAY MACHINE AT THE RADIOLOGY INSTALLATION OF ARIFIN ACHMAD HOSPITAL, RIAU. Medical Imaging and Radiation Protection Research (MIROR) Journal. 2021. doi:10.54973/miror.v1i1.59

BibTeX
@article{Saputra2021,
  author = {Saputra, Yoga and Bisra, Marido},
  title = {CONFORMITY TEST OF THE COLLIMATOR BEAM TO X-RAY ON DIAGNOSTIC X-RAY MACHINE AT THE RADIOLOGY INSTALLATION OF ARIFIN ACHMAD HOSPITAL, RIAU},
  journal = {Medical Imaging and Radiation Protection Research (MIROR) Journal},
  year = {2021},
  doi = {10.54973/miror.v1i1.59}
}
Kostenlos starten →