Comparison of Methods and Verification of Some Performance Characteristics of Accu-Chek® Instant Glucose Meter

KENNETH INAKU, LAWSON EKPE, Asuquo Ene, ONUCHE Onuche, ESSIENAWAN AKPAN, LUCY INAKU

Cross River Journal of Medicine · 2024

DOI: 10.5455/crjmed.175560 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Cross River Journal of Medicine, 2024. Verfasst von KENNETH INAKU und 5 weiteren Autor:innen. Die Arbeit wurde bisher 2-mal zitiert (Crossref).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Frequent and accurate blood glucose measurements and control can delay the onset of diabetic complications. This is facilitated by the use of point-of-care glucose meters (POCGM). There is a need to verify some claims made by manufacturers of POCGM with regard to accuracy, precision and linearity. We compared blood glucose results obtained from Accu-Check Instant blood glucose meters with laboratory results as reference in order to verify some of these claims.

Method

This prospective analytical study involved the verification of some characteristics of Accu-Chek Instant POCGM using different concentrations of glucose calibrators prepared in-house. Blood was obtained from volunteers who had no need for an overnight fast. It was tested using Accu-Chek Instant POCGM and routine laboratory glucose oxidase method as reference and compared using the Bland-Altman plot. Statistical significance was set at p

Zitieren

APA

INAKU, K., EKPE, L., Ene, A., Onuche, O., AKPAN, E., & INAKU, L. (2024). Comparison of Methods and Verification of Some Performance Characteristics of Accu-Chek® Instant Glucose Meter. Cross River Journal of Medicine. https://doi.org/10.5455/crjmed.175560

Vancouver

INAKU K, EKPE L, Ene A, Onuche O, AKPAN E, INAKU L. Comparison of Methods and Verification of Some Performance Characteristics of Accu-Chek® Instant Glucose Meter. Cross River Journal of Medicine. 2024. doi:10.5455/crjmed.175560

BibTeX
@article{INAKU2024,
  author = {INAKU, KENNETH and EKPE, LAWSON and Ene, Asuquo and Onuche, ONUCHE and AKPAN, ESSIENAWAN and INAKU, LUCY},
  title = {Comparison of Methods and Verification of Some Performance Characteristics of Accu-Chek® Instant Glucose Meter},
  journal = {Cross River Journal of Medicine},
  year = {2024},
  doi = {10.5455/crjmed.175560}
}
Kostenlos starten →