Comparative study on the determination of free formaldehyde residue in vaccines by pre-column derivative liquid chromatography and post-column derivatization liquid chromatography

Yukun Chen, Congcong Liu, Weiyang Liang

Advances in Analytic Science · 2023

DOI: 10.54517/aas.v4i2.2009 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Advances in Analytic Science, 2023. Verfasst von Yukun Chen und 2 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To develop and compare methods for quantifying residual free formaldehyde in vaccines using pre-column and post-column derivatization liquid chromatography, and to assess the consistency between these methods.

Methods

For pre-column derivatization, we used a Shimadzu LC-20AT liquid chromatograph equipped with an SPD-20A UV detector. Separation was achieved with a mobile phase consisting of 60% acetonitrile at a flow rate of 0. 8 mL·min−1 and a column temperature of 40 ℃. Detection was performed at 360 nm using a Kromasil 100-5-C18 (250 mm × 4.6 mm) column. For post-column derivatization, we employed a Shimadzu LC-20AT liquid chromatograph with an SPD-M20A diode array detector and a vector derivative device. Separation was conducted with a mobile phase of 0.2% (V/V) phosphoric acid at a flow rate of 1.0 mL·min−1 and a column temperature of 25 ℃. Detection was carried out at 412 nm using a Chrom Core AQ-C18 (250 mm × 4.6 mm) column. The derivatization solution was acetate buffer, flowing at 0.5 mL·min−1 and maintained at 100 ℃. Both methods were evaluated for precision, repeatability, and sample recovery, with statistical significance assessed using F-test and t-test.

Results

The pre-column derivatization method exhibited linearity over the range of 0.025–100 μg·mL−1 (R = 0.9999, n = 12). Precision showed an RSD value of 0.06%, and repeatability ranged from 0.3% to 1.4%. Average recoveries were between 97.3% and 104.8%, with RSD values from 0.7% to 2.9%. The limit of quantitation (LOQ) was 0.02 μg·mL−1, and the limit of detection (LOD) was 0.01 μg·mL−1. The post-column derivatization method demonstrated linearity in the same range (0.025–100 μg·mL−1, R = 0.9999, n = 12). Precision showed an RSD value of 0.02%, and repeatability ranged from 0.7% to 3.5%. Average recoveries were between 105.6% and 114.6%, with RSD values from 0.3% to 1.9%. The LOQ was 0.02 μg·mL−1, and the LOD was 0.006 μg·mL−1. Statistical analysis indicated that both methods produced comparable results.

Conclusion

Both pre-column and post-column derivatization liquid chromatography methods are simple, accurate, and highly sensitive, making them suitable for determining residual free formaldehyde in vaccines.

Zitieren

APA

Chen, Y., Liu, C., & Liang, W. (2023). Comparative study on the determination of free formaldehyde residue in vaccines by pre-column derivative liquid chromatography and post-column derivatization liquid chromatography. Advances in Analytic Science. https://doi.org/10.54517/aas.v4i2.2009

Vancouver

Chen Y, Liu C, Liang W. Comparative study on the determination of free formaldehyde residue in vaccines by pre-column derivative liquid chromatography and post-column derivatization liquid chromatography. Advances in Analytic Science. 2023. doi:10.54517/aas.v4i2.2009

BibTeX
@article{Chen2023,
  author = {Chen, Yukun and Liu, Congcong and Liang, Weiyang},
  title = {Comparative study on the determination of free formaldehyde residue in vaccines by pre-column derivative liquid chromatography and post-column derivatization liquid chromatography},
  journal = {Advances in Analytic Science},
  year = {2023},
  doi = {10.54517/aas.v4i2.2009}
}
Kostenlos starten →