EUVIMEDEUVIMED entdecken
Crossref

Prüfung der Tränenfilmmessungen eines multifunktionalen Instrumentes

Heiko Pult

Optometry & Contact Lenses · 2022

DOI: 10.54352/dozv.bpxm2142 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Optometry & Contact Lenses, 2022. Verfasst von Heiko Pult.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The aim of this study was to test the measurements of the tear film stability and tear meniscus height of the Visio- nix 120+ (VX; Luneau Technology, Jerusalem, Israel). Material and Methods. In 24 volunteers (mean age = 37.42 ± 14.23 (standard deviation)), the tear film stability (V-NIBUT and V-NIBUT M) and the tear meniscus height (V-TMH) were determined with VX and the gold-standard, the “Tearscope” technique, of the right eyes (control; NIBUT and TMH). The symptoms were observed using the Ocular Disease Index-6 (OSDI-6). The methods were analysed with Cohen’s Kappa and intra-class correlation (ICC), the prediction probability with the Receiver-Operator-Characteristic-Curve (ROC).

Results

The average OSDI-6 grade was 8.00 ± 5.75, the average tear film stability, measured with the VX, was 3.92 s ± 1.79 (V-NIBUT) and 7.36 s ± 1.67 (V-NIBUT M), the control was 10.57 s ± 6.84. The mean tear meniscus height, measured with the VX, was 0.31 mm ± 0.10, the control 0.31 mm ± 0.11. The Cohen’s Kappa analysis showed none to slight agree- ment between methods (V-NIBUT - NIBUT: Kappa -0.005, p = 0.708; V-NIBUT M – NIBUT: Kappa 0.035, p = 0.032; V-TMH - TMH: Kappa 0.06, p = 0.051). V-NIBUT to NIBUT gave an ICC of 0.167 (p = 0.332), V-NIBUT M to NIBUT gave 0.296 (p = 0,203) and V-TMH to TMH -0.192 (p = 0.661). While NIBUT was a significant discriminator of OSDI-6, this was not the case with V-NIBUT or V-NIBUT M (NIBUT: area under the ROC (AUC) = 0.832, p = 0.009; V-NIBUT: AUC = 0.547; p = 0.145; V-NIBUT M: AUC = 0.668, p = 0.127).

Conclusion

The measurements of the VX differed signifi- cantly from the control measurements. Keywords tear film stability, NIBUT, tear meniscus height, OSDI-6, dry eye

Zitieren

APA

Pult, H. (2022). Prüfung der Tränenfilmmessungen eines multifunktionalen Instrumentes. Optometry & Contact Lenses. https://doi.org/10.54352/dozv.bpxm2142

Vancouver

Pult H. Prüfung der Tränenfilmmessungen eines multifunktionalen Instrumentes. Optometry & Contact Lenses. 2022. doi:10.54352/dozv.bpxm2142

BibTeX
@article{Pult2022,
  author = {Pult, Heiko},
  title = {Prüfung der Tränenfilmmessungen eines multifunktionalen Instrumentes},
  journal = {Optometry & Contact Lenses},
  year = {2022},
  doi = {10.54352/dozv.bpxm2142}
}
Kostenlos starten →