Inappropriate Asystole Detection in Early Postoperative Phase after Loop Recorder Implantation
ISRN Cardiology · 2011
DOI: 10.5402/2011/146062 ↗Kurz gefasst
Fallbericht in der Zeitschrift ISRN Cardiology, 2011. Verfasst von Miriam Bortnik und 4 weiteren Autor:innen. Beteiligte Einrichtung: Cardiology Division, Azienda Ospedaliera Maggiore della Carità, 28100 Novara, Italy. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 10 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 3.0).
Abstract (Originalsprache)
The implantable loop recorder is a useful diagnostic tool for patients with unexplained syncope. The capability to automatically detect and store arrhythmic events, implemented in the last generations of these devices, can further improve the diagnostic yield, but this feature can be compromised by inappropriate detection of false arrhythmias. We herein report the case of a patient in which several inappropriate activations of long-lasting asystole occurred in the two days following the implant, probably because of an intermittently loose contact between the device and subcutaneous tissue for a small pocket haematoma.
Autor:innen und Institutionen
- Cardiology Division, Azienda Ospedaliera Maggiore della Carità, 28100 Novara, Italy.
Zitieren
APA
Bortnik, M., Occhetta, E., Magnani, A., Degiovanni, A., & Marino, P. (2011). Inappropriate Asystole Detection in Early Postoperative Phase after Loop Recorder Implantation. ISRN Cardiology. https://doi.org/10.5402/2011/146062
Vancouver
Bortnik M, Occhetta E, Magnani A, Degiovanni A, Marino P. Inappropriate Asystole Detection in Early Postoperative Phase after Loop Recorder Implantation. ISRN Cardiology. 2011. doi:10.5402/2011/146062
BibTeX
@article{Bortnik2011,
author = {Bortnik, Miriam and Occhetta, Eraldo and Magnani, Andrea and Degiovanni, Anna and Marino, Paolo},
title = {Inappropriate Asystole Detection in Early Postoperative Phase after Loop Recorder Implantation},
journal = {ISRN Cardiology},
year = {2011},
doi = {10.5402/2011/146062}
}
EUVIMED