JAMINAN KUALITAS EVALUASI DOSIS HP(10) DAN HP(0,07) PEKERJA RADIASI DI SEKTOR KESEHATAN
Prosiding Seminar Si-INTAN · 2022
DOI: 10.53862/ssi.v2.072022.006 ↗Kurz gefasst
Fachartikel in der Zeitschrift Prosiding Seminar Si-INTAN, 2022. Verfasst von Nuha Nabilah Utrujjah und 2 weiteren Autor:innen.
Abstract (Originalsprache)
As one of the externa laboratories engaged in radiation safety, NuklindoLab has an individual dose evaluation service hp (10) and hp (0.07) using a TLD CaSO4:Dy thermoluminiscence dosimeter. Currently NuklindoLab has served as many as 1350 customers in the health sector consisting of clinics and hospitals classes A, B, and C, so that dose evaluation services must be guaranteed well. This study covers the internal and external quality assurance of Hp(10) and Hp(0.07) dose evaluation services at NuklindoLab – Kop JKRL. Internal quality assurance is carried out, namely background analysis of TLD CaSO4:Dy, QC Chart TLD Reader, calibration curve, all of which are in a good range. To verify this, calibration was carried out with a mixed radiation field Cs-137 and Sr-90 with dose variations of 10x, 15x, and 20x of the detection limit. The resulting uncertainty is still below 42% so it is still within the range of the IAEA tolerance limit. In addition, blind tests were also carried out to see the performance of equipment and personnel with mixed irradiation between Cs-137 and Sr-90 and with the ash of gas mantles. The deviation obtained by the ash of gas mantles is better, which is 11.2% for Hp (10) and 10.8% for Hp (0.07) so this indicates that the blind test can also be done with the source ash of gas mantles. Meanwhile, the external quality assurance carried out, namely comparing the external intercomparation that have been carried out in 2019 and 2021, has good performance, where the results obtained are 95% in the trumpet curve. Keywords: quality assurance, individual dosimeter, TLD CaSO4:Dy, beta radiation, gama radiation
Zitieren
APA
Utrujjah, N. N., Yuana, F., & Bunawas, B. (2022). JAMINAN KUALITAS EVALUASI DOSIS HP(10) DAN HP(0,07) PEKERJA RADIASI DI SEKTOR KESEHATAN. Prosiding Seminar Si-INTAN. https://doi.org/10.53862/ssi.v2.072022.006
Vancouver
Utrujjah NN, Yuana F, Bunawas B. JAMINAN KUALITAS EVALUASI DOSIS HP(10) DAN HP(0,07) PEKERJA RADIASI DI SEKTOR KESEHATAN. Prosiding Seminar Si-INTAN. 2022. doi:10.53862/ssi.v2.072022.006
BibTeX
@article{Utrujjah2022,
author = {Utrujjah, Nuha Nabilah and Yuana, Firdy and Bunawas, Bunawas},
title = {JAMINAN KUALITAS EVALUASI DOSIS HP(10) DAN HP(0,07) PEKERJA RADIASI DI SEKTOR KESEHATAN},
journal = {Prosiding Seminar Si-INTAN},
year = {2022},
doi = {10.53862/ssi.v2.072022.006}
}
EUVIMED