A theoretical framework for esophageal pressure driven closed-loop mechanical ventilation mode

Ehab Daoud

Journal of Mechanical Ventilation · 15.06.2026 · 7 (2) · 78-86

DOI: 10.53097/jmv.10153 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Mechanical Ventilation, online erschienen am 15.06.2026. Verfasst von Ehab Daoud. Die Arbeit nennt 6 Literaturangaben.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Esophageal balloon manometry is a surrogate for pleural pressure and has been in clinical use for decades. The main advantages for its use are to partition the total lung and chest wall mechanics, thus providing transpulmonary inspiratory, expiratory, and driving pressure monitoring as the true stress delivered to the lung, and the capability to measure the transpulmonary mechanical power as the true energy delivered to inflate the lungs. Thus, it is able to provide more insight into lung-protective ventilation than total airway pressure alone. Its capability as the gold standard to measure patient effort, muscle pressure, and to diagnose patient-ventilator interactions and dyssynchronies adds further value to its usage. Proposed Framework As the paradigm shifts toward personalized lung and diaphragm-protective ventilation, alongside increased recognition of ventilator-induced and patient self-inflicted lung injury (VILI and P-SILI respectively), using esophageal balloon manometry seems to be the ultimate tool to achieve these goals. In this paper, we introduce a hypothetical automated closed-loop ventilator mode that is based on adjusting the inspiratory and expiratory transpulmonary pressures to a safe zone to provide true lung-protective ventilation, while simultaneously adjusting the triggering, maintaining, and cycling of breaths according to the patient's neural time and effort, potentially eliminating dyssynchronies. Clinical Implications This adaptive mode aims to optimize the balance between lung recruitment, overdistension, and respiratory muscle unloading, addressing VILI, P-SILI and ventilator-induced diaphragm dysfunction (VIDD).

Conclusion

A new closed-loop ventilator mode based on continuous input and feedback from the esophageal balloon can conceptually lead to safer personalized ventilation, significantly lessen dyssynchronies, and potentially improve mortality outcomes in acute respiratory failure. Keywords: Esophageal balloon manometry, Transpulmonary pressures, dyssynchronies, closed-loop ventilation.

Zitieren

APA

Daoud, E. (2026). A theoretical framework for esophageal pressure driven closed-loop mechanical ventilation mode. Journal of Mechanical Ventilation, 7(2), 78-86. https://doi.org/10.53097/jmv.10153

Vancouver

Daoud E. A theoretical framework for esophageal pressure driven closed-loop mechanical ventilation mode. Journal of Mechanical Ventilation. 2026;7(2):78-86. doi:10.53097/jmv.10153

BibTeX
@article{Daoud2026,
  author = {Daoud, Ehab},
  title = {A theoretical framework for esophageal pressure driven closed-loop mechanical ventilation mode},
  journal = {Journal of Mechanical Ventilation},
  year = {2026},
  volume = {7},
  number = {2},
  pages = {78--86},
  doi = {10.53097/jmv.10153}
}
Kostenlos starten →