Kurz gefasst

Fachartikel in der Zeitschrift International Neurourology Journal, erschienen am 31.05.2023. Verfasst von Byeong-Il Kang und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Beckman Laser Institute Korea, School of Medicine, Dankook University, Cheonan, Korea. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 3.0). Gefördert von: National Research Foundation of Korea, Ministry of Science and Technology, Ministry of Education.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Current guidelines recommend clean intermittent catheterization (CIC) at regular time intervals for patients with spinal cord injuries; however, many patients experience difficulties. Performing time-based CIC outside the home is a significant burden for patients. In this study, we aimed to overcome the limitations of the current guidelines by developing a digital device to monitor bladder urine volume in real-time.

Methods

The optode sensor is a near-infrared spectroscopy (NIRS)-based wearable device intended to be attached to the skin of the lower abdomen where the bladder is located. The sensor’s primary function is to detect changes in urine volume within the bladder. An in vitro study was conducted using a bladder phantom that mimicked the optical properties of the lower abdomen. To validate the data in the human body at the proof-of-concept level, one volunteer attached the device to the lower abdomen to measure the light intensity between the first voiding and immediately before the second voiding.

Results

The degree of attenuation at the maximum test volume was equivalent across experiments, and the optode sensor with multiplex measurements demonstrated robust performance for patient diversity. Moreover, the symmetric feature of the matrix was deemed a potential parameter for identifying the accuracy of sensor localization in a deep-learning model. The validated feasibility of the sensor showed almost the same results as an ultrasound scanner, which is routinely used in the clinical field.

Conclusions

The optode sensor of the NIRS-based wearable device can measure the urine volume in the bladder in real-time.

Schlagwörter

Stichwörter der Autor:innen

Bladder monitoringClean intermittent catheterizationNeurogenic bladder

Autor:innen und Institutionen

  • Department of Biomedical Engineering, Beckman Laser Institute Korea, School of Medicine, Dankook University, Cheonan, Korea.
  • MEDiThings Co., Ltd., Seoul, Korea.
  • Department of Urology and Neurogenic Bladder Clinic, Konkuk University Medical Center, Konkuk University School of Medicine, Seoul, Korea.

Förderung

  • National Research Foundation of Korea
  • Ministry of Science and Technology
  • Ministry of Education
  • Korea Technology Information Promotion Agency
  • Ministry of SMEs and Startups

Zitieren

APA

Kang, B. I., Kim, A., & Kim, S. (2023). Advancing Patient Care: Innovative Use of Near-Infrared Spectroscopy for Monitoring Urine Volume in Neurogenic Bladder. International Neurourology Journal, 27(Suppl 1), S27-33. https://doi.org/10.5213/inj.2346100.050

Vancouver

Kang BI, Kim A, Kim S. Advancing Patient Care: Innovative Use of Near-Infrared Spectroscopy for Monitoring Urine Volume in Neurogenic Bladder. International Neurourology Journal. 2023;27(Suppl 1):S27-33. doi:10.5213/inj.2346100.050

BibTeX
@article{Kang2023,
  author = {Kang, Byeong-Il and Kim, Aram and Kim, Sehwan},
  title = {Advancing Patient Care: Innovative Use of Near-Infrared Spectroscopy for Monitoring Urine Volume in Neurogenic Bladder},
  journal = {International Neurourology Journal},
  year = {2023},
  volume = {27},
  number = {Suppl 1},
  pages = {S27--33},
  doi = {10.5213/inj.2346100.050}
}
Kostenlos starten →