New electronic apex locator Romiapex A-15 presented accuracy for working length determination in permanent teeth

Etevaldo Matos Maia Filho, Cláudia de Castro Rizzi, Daniela Silva Barroso de Oliveira, Paulo Nelson Filho, Raquel Assed Bezerra da Silva, Léa Assed Bezerra da Silva

Dentistry 3000 · 2014

DOI: 10.5195/d3000.2014.26 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Dentistry 3000, 2014. Verfasst von Etevaldo Matos Maia Filho und 5 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The present study aims to evaluate, ex vivo, the accuracy of electronic apex locators Root ZX II and Romiapex-15, for working length (WL) determination in permanent teeth. Materials and Methods: Fourteen single-rooted teeth (incisors and canines), with their apices fully formed were used. The dental crowns were removed. The anatomic length of the tooth (real measurement) was visually determined through the insertion of a size 10 K-file until the tip of the instrument could be observed in the apical foramen under a microscope (20X). Teeth were fixed in a model of resin and adapted into alginate soaked with saline solution, which was used as an electrical conductor. Using a K-file, root canals were measured electronically using both devices. The results obtained for each apex locator were compared to the real measurements. The accuracy between the devices was statistically analyzed using the Bland-Altman graph, Intraclass Correlation Coefficient (ICC), and Student’s t-test.

Results

The mean difference between the measurements using the Root ZX II was 0.277mm greater than the real measurement, while the measurements from the Romiapex-15 were 0.308mm higher on average. The comparison between Root ZX II and Romiapex-15 had no significant difference (p= 0.868).

Conclusion

It was concluded that Root ZX II and Romiapex-15 had similar accuracy. Romiapex-15 could be an option for WL determination in permanent teeth.

Zitieren

APA

Filho, E. M. M., Rizzi, C. D. C., Oliveira, D. S. B. D., Filho, P. N., Silva, R. A. B. D., & Silva, L. A. B. D. (2014). New electronic apex locator Romiapex A-15 presented accuracy for working length determination in permanent teeth. Dentistry 3000. https://doi.org/10.5195/d3000.2014.26

Vancouver

Filho EMM, Rizzi CDC, Oliveira DSBD, Filho PN, Silva RABD, Silva LABD. New electronic apex locator Romiapex A-15 presented accuracy for working length determination in permanent teeth. Dentistry 3000. 2014. doi:10.5195/d3000.2014.26

BibTeX
@article{Filho2014,
  author = {Filho, Etevaldo Matos Maia and Rizzi, Cláudia de Castro and Oliveira, Daniela Silva Barroso de and Filho, Paulo Nelson and Silva, Raquel Assed Bezerra da and Silva, Léa Assed Bezerra da},
  title = {New electronic apex locator Romiapex A-15 presented accuracy for working length determination in permanent teeth},
  journal = {Dentistry 3000},
  year = {2014},
  doi = {10.5195/d3000.2014.26}
}
Kostenlos starten →