Biomechanical Impact of Implant Diameter and Bone Quality on Single-Crown Restorations: A Finite Element Study
International Journal of Dental Research and Allied Sciences · 2021
DOI: 10.51847/58ewtgx0xs ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Dental Research and Allied Sciences, 2021. Verfasst von Juan Rodríguez und 2 weiteren Autor:innen. Die Arbeit nennt 31 Literaturangaben.
Abstract (Originalsprache)
The performance of implant-supported restorations is strongly influenced by both the implant diameter and the surrounding bone density. This research aimed to examine how these two parameters affect stress distribution when subjected to vertical and 30° oblique loading. A finite element analysis was performed using dental implants of a constant length of 6.5 mm, but with different diameters (Ø3.3, Ø3.5, Ø3.75, Ø4.0, Ø4.25, and Ø4.75 mm). Each implant was positioned axially and restored using a 2
Zitieren
APA
Rodríguez, J., Stewart, Z., & González, M. (2021). Biomechanical Impact of Implant Diameter and Bone Quality on Single-Crown Restorations: A Finite Element Study. International Journal of Dental Research and Allied Sciences. https://doi.org/10.51847/58ewtgx0xs
Vancouver
Rodríguez J, Stewart Z, González M. Biomechanical Impact of Implant Diameter and Bone Quality on Single-Crown Restorations: A Finite Element Study. International Journal of Dental Research and Allied Sciences. 2021. doi:10.51847/58ewtgx0xs
BibTeX
@article{Rodrguez2021,
author = {Rodríguez, Juan and Stewart, Zoey and González, María},
title = {Biomechanical Impact of Implant Diameter and Bone Quality on Single-Crown Restorations: A Finite Element Study},
journal = {International Journal of Dental Research and Allied Sciences},
year = {2021},
doi = {10.51847/58ewtgx0xs}
}
EUVIMED