In Vivo Bone Mineral Density Assessment With Spectral Localizer Radiographs From Photon-Counting Detector CT: Prospective Comparison With DXA

Moser, Lukas Jakob, Klambauer, Konstantin, Diaz Machicado, Maria Carolina, Frey, Diana, Mergen, Victor, Eberhard, Matthias und 10 weitere

Zurich Open Repository and Archive (University of Zurich) · 19.02.2025

DOI: 10.5167/uzh-275513 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Zurich Open Repository and Archive (University of Zurich), erschienen am 19.02.2025. Verfasst von Moser und 21 weiteren Autor:innen.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE The aim of this study was to determine in a prospective patient study the accuracy of areal bone mineral density (aBMD) measurements with spectral localizer radiographs obtained with a clinical photon-counting detector computed tomography (PCD-CT) scanner in comparison with dual-energy x-ray absorptiometry (DXA). METHODS In this institutional review board-approved, prospective study, 41 patients (15 females, 26 males; mean age 61.3 years, age range 35-78 years) underwent PCD-CT of the abdomen with a spectral localizer radiograph (tube voltage 140 kVp, tube current 30 mA) and DXA within a median of 45 days. aBMD values were derived for lumbar vertebrae L1-L4 from both methods and were compared with linear regression, Pearson correlation, intraclass correlation coefficients (ICCs), and Bland-Altman plots. T-scores were calculated on a patient level and were compared between methods. RESULTS DXA and spectral localizer radiographs showed strong correlation in aBMD measurements (R = 0.97, P < 0.001) and patient level T-scores (R = 0.99, P < 0.001). There was a strong agreement between aBMD from both methods (ICC, 0.96; 95% CI, 0.94-0.97). Bland-Altman analysis revealed a very small mean difference in aBMD between methods (mean absolute error 0.019 g/cm2) with narrow limits of agreement (-0.083 g/cm2 to 0.121 g/cm2). Similarly, there were small differences in regard to the T-score (mean absolute error 0.156) with narrow limits of agreement (-0.422 to 0.734) between methods. ICCs indicated an excellent agreement between T-scores from DXA and spectral localizer radiographs (ICC, 0.98; 95% confidence interval, 0.95-0.99). CONCLUSIONS Our prospective patient study indicates that spectral localizer radiographs obtained with a clinical PCD-CT system enable accurate quantification of the lumbar bone areal mineral density. This opens up the opportunity for opportunistic screening of osteoporosis in patients who undergo CT for other indications.

Schlagwörter

bone mineral densityphoton-counting CTdual-energy X-ray absorptiometrylumbar vertebraeosteoporosis screeningT-scoremeasurement agreementlow-dose CTopportunistic screeningosteoporosis

Zitieren

APA

Moser, Jakob, L., Klambauer, Konstantin, Machicado, D., Carolina, M., Frey, Diana, Mergen, Victor, Eberhard, Matthias, Nowak, Tristan, Schmidt, Bernhard, Flohr, Thomas, Distler, …, & Hatem (2025). In Vivo Bone Mineral Density Assessment With Spectral Localizer Radiographs From Photon-Counting Detector CT: Prospective Comparison With DXA. Zurich Open Repository and Archive (University of Zurich). https://doi.org/10.5167/uzh-275513

Vancouver

Moser, Jakob L, Klambauer, Konstantin, Machicado D, Carolina M, et al. In Vivo Bone Mineral Density Assessment With Spectral Localizer Radiographs From Photon-Counting Detector CT: Prospective Comparison With DXA. Zurich Open Repository and Archive (University of Zurich). 2025. doi:10.5167/uzh-275513

BibTeX
@article{Moser2025,
  author = {Moser and Jakob, Lukas and Klambauer and Konstantin and Machicado, Diaz and Carolina, Maria and Frey and Diana and Mergen and Victor and Eberhard and Matthias and Nowak and Tristan and Schmidt and Bernhard and Flohr and Thomas and Distler and Oliver and Alkadhi and Hatem},
  title = {In Vivo Bone Mineral Density Assessment With Spectral Localizer Radiographs From Photon-Counting Detector CT: Prospective Comparison With DXA},
  journal = {Zurich Open Repository and Archive (University of Zurich)},
  year = {2025},
  doi = {10.5167/uzh-275513}
}
Kostenlos starten →