Kurz gefasst

Fallbericht in der Zeitschrift Journal of Education and Teaching in Emergency Medicine, online erschienen am 31.10.2026. Verfasst von Clifford J Buckley und 3 weiteren Autor:innen. Beteiligte Einrichtung: Baylor Scott & White Medical Center, Department of Emergency Medicine, Temple, TX. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Synchronized cardioversion is a standard treatment for unstable ventricular tachycardia. While modern defibrillators are designed with safety mechanisms, synchronization failure can result in catastrophic complications. We present the case of a 65-year-old woman with unstable wide-complex tachycardia with a pulse who underwent emergent synchronized cardioversion. Despite activation of the “sync” mode, the device failed to capture appropriately and delivered a shock on the T-wave, precipitating ventricular fibrillation and cardiac arrest. After 26 minutes of resuscitation and multiple defibrillations, return of spontaneous circulation (ROSC) was achieved. The patient ultimately recovered and was discharged home in stable condition. This case underscores the rare but serious risk of synchronization failure, particularly in obese patients where pad placement and capture may be impaired,1 and emphasizes the importance of visually confirming synchronization prior to shock delivery.

Autor:innen und Institutionen

  • Baylor Scott & White Medical Center, Department of Emergency Medicine, Temple, TX.
  • Baylor College of Medicine, Temple, TX.

Zitieren

APA

Buckley, C. J., McClintic, B. J., Juergens, A. L., & Standley, S. (2026). Shockingly Unreliable: A Case Report of Failure of Synchronized Cardioversion Device to Capture, Resulting in Ventricular Fibrillation and Cardiac Arrest. Journal of Education and Teaching in Emergency Medicine, 11(4). https://doi.org/10.5070/m5.53094

Vancouver

Buckley CJ, McClintic BJ, Juergens AL, Standley S. Shockingly Unreliable: A Case Report of Failure of Synchronized Cardioversion Device to Capture, Resulting in Ventricular Fibrillation and Cardiac Arrest. Journal of Education and Teaching in Emergency Medicine. 2026;11(4). doi:10.5070/m5.53094

BibTeX
@article{Buckley2026,
  author = {Buckley, Clifford J and McClintic, Bradley J and Juergens, Andrew L and Standley, Shea},
  title = {Shockingly Unreliable: A Case Report of Failure of Synchronized Cardioversion Device to Capture, Resulting in Ventricular Fibrillation and Cardiac Arrest},
  journal = {Journal of Education and Teaching in Emergency Medicine},
  year = {2026},
  volume = {11},
  number = {4},
  doi = {10.5070/m5.53094}
}
Kostenlos starten →