Comparative Study of MiroCam MC2000 and PillCam SB3 in Detecting Small Bowel Bleeding: A Multicenter Prospective Randomized Crossover Study
Gut and Liver · 2025
DOI: 10.5009/gnl240541 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Gut and Liver, 2025. Verfasst von Ji Eun Kim und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea. Hauptthemen laut MEDLINE: Gastrointestinal Hemorrhage, Intestine, Small, Capsule Endoscopy. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Ministry of Science and ICT, South Korea, Ministry of Trade, Industry and Energy, Ministry of Health and Welfare.
Abstract (Originalsprache)
Background/Aims: The MiroCam MC2000 (MC2000) is a double-tip capsule with a camera on each side. It is designed to provide more extensive visualization of the small bowel mucosa, potentially reducing the chance of missing lesions. This study aimed to compare the detection rates for lesions in the ampulla of Vater (AoV) and the small bowel of the MC2000 and the PillCam SB3 (SB3) for patients with suspected small bowel bleeding.
Methods
This prospective, multicenter, randomized crossover trial compared the lesion detection capabilities of the MC2000 and SB3 capsules, ingested one hour apart by patients with suspected small bowel bleeding. The primary outcome was the detection of lesions in the AoV, while the secondary outcome was the assessment of the detection of P1 and P2 lesions, known underlying causes of small bowel bleeding.
Results
There was no significant difference in AoV lesion detection rates between the devices. However, MC2000 demonstrated significantly greater detection of red spots in patients with visible bleeding (p=0.018) and tended to detect a greater number of small bowel lesions, including P2 lesions. Minor complications included device stasis, with fewer incidents with the MC2000 than with the SB3, and one instance of small bowel retention due to ulcers.
Conclusions
The MC2000's dual-camera system appears to enhance the detection of small bowel lesions over the SB3, especially for more important lesions. These findings suggest that the MC2000 may offer superior diagnostic capabilities for patients with suspected small bowel bleeding, potentially leading to better clinical outcomes (this trial registered KCT0005591).
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea.
- Division of Gastroenterology, Department of Internal Medicine, Severance Hospital, Yonsei University College of Medicine, Seoul, Korea.
- Department of Internal Medicine, Yeungnam University College of Medicine, Daegu, Korea.
- Department of Internal Medicine, Ewha Womans University Seoul Hospital, Seoul, Korea.
- Department of Internal Medicine, Hallym University College of Medicine, Chuncheon, Korea.
- Department of Medicine, Hallym University Dongtan Sacred Heart Hospital, Hallym University College of Medicine, Hwaseong, Korea.
ORCID-Profile: Ji Eun Kim, Eun Ran Kim, Jae Jun Park, Kyeong Ok Kim, Yehyun Park, Young Joo Yang, Hyun Joo Jang
Förderung
- Ministry of Science and ICT, South Korea
- Ministry of Trade, Industry and Energy
- Ministry of Health and Welfare
- Ministry of Food and Drug Safety
Zitieren
APA
Kim, J. E., Kim, E. R., Park, J. J., Kim, K. O., Park, Y., Yang, Y. J., & Jang, H. J. (2025). Comparative Study of MiroCam MC2000 and PillCam SB3 in Detecting Small Bowel Bleeding: A Multicenter Prospective Randomized Crossover Study. Gut and Liver. https://doi.org/10.5009/gnl240541
Vancouver
Kim JE, Kim ER, Park JJ, Kim KO, Park Y, Yang YJ, et al. Comparative Study of MiroCam MC2000 and PillCam SB3 in Detecting Small Bowel Bleeding: A Multicenter Prospective Randomized Crossover Study. Gut and Liver. 2025. doi:10.5009/gnl240541
BibTeX
@article{Kim2025,
author = {Kim, Ji Eun and Kim, Eun Ran and Park, Jae Jun and Kim, Kyeong Ok and Park, Yehyun and Yang, Young Joo and Jang, Hyun Joo},
title = {Comparative Study of MiroCam MC2000 and PillCam SB3 in Detecting Small Bowel Bleeding: A Multicenter Prospective Randomized Crossover Study},
journal = {Gut and Liver},
year = {2025},
doi = {10.5009/gnl240541}
}
EUVIMED