A Case of Spuriously Decreased White Blood Cell Count on an Automated Sysmex XN Hematology Analyzer: The Difference Between the WNR and WDF Channels

Han Joo Kim, Seung Namgoong, M.T., Chan-Jeoung Park

Laboratory Medicine Online · 01.07.2021 · 11 (3) · 191-193

DOI: 10.47429/lmo.2021.11.3.191 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Laboratory Medicine Online, erschienen am 01.07.2021. Verfasst von Han Joo Kim und 3 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 6 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

be accurate and precise.However, in some cases, spurious CBC can occur, which may require additional action on the part of the clinician to obtain correct results [1,2].We report a case of an incorrect white blood cell (WBC) count obtained on a Sysmex XN hematology analyzer (Sysmex Corporation, Kobe, Japan). CASE REPORTA 56-year-old man visited our hospital for follow-up of hepatocellular carcinoma with lung metastases.He had started sorafenib (a molecular targeted multi-kinase inhibitor) 3 months before and nivolumab (a humanized monoclonal antibody to the immune checkpoint receptor programmed death 1 [PD-1]) 2 days before his peripheral blood was drawn for laboratory tests, including CBC.The initial WBC count measured on a Sysmex XN hematology analyzer (Sysmex Corporation) was 0.02×10 3 /μL; the differential count was not available due to the low WBC count.His he-

Zitieren

APA

Kim, H. J., Namgoong, S., M.T., & Park, C. (2021). A Case of Spuriously Decreased White Blood Cell Count on an Automated Sysmex XN Hematology Analyzer: The Difference Between the WNR and WDF Channels. Laboratory Medicine Online, 11(3), 191-193. https://doi.org/10.47429/lmo.2021.11.3.191

Vancouver

Kim HJ, Namgoong S, M.T., Park C. A Case of Spuriously Decreased White Blood Cell Count on an Automated Sysmex XN Hematology Analyzer: The Difference Between the WNR and WDF Channels. Laboratory Medicine Online. 2021;11(3):191-193. doi:10.47429/lmo.2021.11.3.191

BibTeX
@article{Kim2021,
  author = {Kim, Han Joo and Namgoong, Seung and M.T. and Park, Chan-Jeoung},
  title = {A Case of Spuriously Decreased White Blood Cell Count on an Automated Sysmex XN Hematology Analyzer: The Difference Between the WNR and WDF Channels},
  journal = {Laboratory Medicine Online},
  year = {2021},
  volume = {11},
  number = {3},
  pages = {191--193},
  doi = {10.47429/lmo.2021.11.3.191}
}
Kostenlos starten →