A Case of Spuriously Decreased White Blood Cell Count on an Automated Sysmex XN Hematology Analyzer: The Difference Between the WNR and WDF Channels
Laboratory Medicine Online · 01.07.2021 · 11 (3) · 191-193
DOI: 10.47429/lmo.2021.11.3.191 ↗Kurz gefasst
Fachartikel in der Zeitschrift Laboratory Medicine Online, erschienen am 01.07.2021. Verfasst von Han Joo Kim und 3 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 6 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
be accurate and precise.However, in some cases, spurious CBC can occur, which may require additional action on the part of the clinician to obtain correct results [1,2].We report a case of an incorrect white blood cell (WBC) count obtained on a Sysmex XN hematology analyzer (Sysmex Corporation, Kobe, Japan). CASE REPORTA 56-year-old man visited our hospital for follow-up of hepatocellular carcinoma with lung metastases.He had started sorafenib (a molecular targeted multi-kinase inhibitor) 3 months before and nivolumab (a humanized monoclonal antibody to the immune checkpoint receptor programmed death 1 [PD-1]) 2 days before his peripheral blood was drawn for laboratory tests, including CBC.The initial WBC count measured on a Sysmex XN hematology analyzer (Sysmex Corporation) was 0.02×10 3 /μL; the differential count was not available due to the low WBC count.His he-
Zitieren
APA
Kim, H. J., Namgoong, S., M.T., & Park, C. (2021). A Case of Spuriously Decreased White Blood Cell Count on an Automated Sysmex XN Hematology Analyzer: The Difference Between the WNR and WDF Channels. Laboratory Medicine Online, 11(3), 191-193. https://doi.org/10.47429/lmo.2021.11.3.191
Vancouver
Kim HJ, Namgoong S, M.T., Park C. A Case of Spuriously Decreased White Blood Cell Count on an Automated Sysmex XN Hematology Analyzer: The Difference Between the WNR and WDF Channels. Laboratory Medicine Online. 2021;11(3):191-193. doi:10.47429/lmo.2021.11.3.191
BibTeX
@article{Kim2021,
author = {Kim, Han Joo and Namgoong, Seung and M.T. and Park, Chan-Jeoung},
title = {A Case of Spuriously Decreased White Blood Cell Count on an Automated Sysmex XN Hematology Analyzer: The Difference Between the WNR and WDF Channels},
journal = {Laboratory Medicine Online},
year = {2021},
volume = {11},
number = {3},
pages = {191--193},
doi = {10.47429/lmo.2021.11.3.191}
}
EUVIMED