Investigation of Current Conduction Levels of Sterilized Peripheral Nerve Block Needles: Invitro Study
IAR Journal of Anaesthesiology and Critical Care · 2021
DOI: 10.47310/iarjacc.2021.02.01.025 ↗Kurz gefasst
Fachartikel in der Zeitschrift IAR Journal of Anaesthesiology and Critical Care, 2021. Verfasst von Eyüp Aydoğan und 1 weiteren Autor:innen.
Abstract (Originalsprache)
Introduction
Medical products are mostly manufactured for single use, including peripheral nerve block needles [1]. Under these conditions, the reuse of disposable peripheral nerve block needles has been thought to be a saving tool. Materials and Method: The Nerve stimulator and multimeter were connected to the peripheral block needles and the control needle, which had been sterilized with the technique described below and used 20 times on the artificial model tissue beforehand, and the stimulator was given current with the stimulator.
Results
When the control needle and 1, 2, 3, 4, 6, 7, 10 needles were compared, in the independent sample t test analysis, the p value was 0.14, and no statistical significance was found. When the control needle and the 5,8,9 needles were compared, the p value was found to be
Zitieren
APA
Aydoğan, E., & Tire, Y. (2021). Investigation of Current Conduction Levels of Sterilized Peripheral Nerve Block Needles: Invitro Study. IAR Journal of Anaesthesiology and Critical Care. https://doi.org/10.47310/iarjacc.2021.02.01.025
Vancouver
Aydoğan E, Tire Y. Investigation of Current Conduction Levels of Sterilized Peripheral Nerve Block Needles: Invitro Study. IAR Journal of Anaesthesiology and Critical Care. 2021. doi:10.47310/iarjacc.2021.02.01.025
BibTeX
@article{Aydoan2021,
author = {Aydoğan, Eyüp and Tire, Yasin},
title = {Investigation of Current Conduction Levels of Sterilized Peripheral Nerve Block Needles: Invitro Study},
journal = {IAR Journal of Anaesthesiology and Critical Care},
year = {2021},
doi = {10.47310/iarjacc.2021.02.01.025}
}
EUVIMED